  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Geisingen",
      "state": "",
      "country": "Germany",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "95770",
      "recalling_firm": "Trokamed GmbH",
      "address_1": "Kleine Breite 17",
      "address_2": "",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "US (Domestic) distribution in state of PA and OUS (foreign) to country of Germany. .",
      "recall_number": "Z-0716-2025",
      "product_description": "Sheath, 18 Fr., 160 mm (REF: WA2PS18L), gastroenterology-urology endoscopic access overtube.",
      "product_quantity": "102 units",
      "reason_for_recall": "If the inflow speed is too quick, it could result in reduced outflow through the shaft due to blood, tissue fragments and kidney stone fragments, and quickly build up of intrarenal pressure.",
      "recall_initiation_date": "20241127",
      "center_classification_date": "20250110",
      "report_date": "20250122",
      "code_info": "REF: WA2PS18L; UDI-DI: 04251303810872; Lot Numbers: 154214",
      "more_code_info": ""
    }
  ]
}