  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Nevele",
      "state": "",
      "country": "Belgium",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "93556",
      "recalling_firm": "mo-Vis BVBA",
      "address_1": "Biebuyckstraat 15D",
      "address_2": "",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US States: TN, TX.",
      "recall_number": "Z-0713-2024",
      "product_description": "Scoot Control R-net, Product Number P015-61. The Scoot Control is intended to support a wheelchair attendant in controlling or maneuvering a power chair",
      "product_quantity": "129 units",
      "reason_for_recall": "In rare cases, the lever of the throttle can get temporarily stuck during the movement and does not come back to its neutral point by itself. If this happens during usage, an unwanted movement of the chair could be the result. An unwanted movement could result in damage to the wheelchair, material damage around the chair and/or injury to the user, attendant or by-standers.",
      "recall_initiation_date": "20231128",
      "center_classification_date": "20240111",
      "report_date": "20240117",
      "code_info": "UDI-DI: 05407008320119; Lot Numbers: 3304-3317 3322-3329 3440-3452 3569-3618 3830-3844 4020-4054 4170-4176 4250-4255 4321-4330",
      "more_code_info": ""
    }
  ]
}