  "results": [
    {
      "status": "Terminated",
      "city": "Malvern",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "79214",
      "recalling_firm": "Siemens Medical Solutions USA, Inc",
      "address_1": "40 Liberty Blvd",
      "address_2": "N/A",
      "postal_code": "19355-1418",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "Distributed in XX states: Ak, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY, the District of Columbia, Guam, and Puerto Rico.",
      "recall_number": "Z-0711-2018",
      "product_description": "Axiom Artis model Interventional fluoroscopic x-ray foot switch accessory; Model Numbers: 4787797, 4787805, and 4787813.",
      "product_quantity": "142",
      "reason_for_recall": "The possibility exists for the wireless foot switch to fail due to impacts by external factors, such as electrostatic discharge that exceed a certain intensity. If the foot switch fails, it will no longer be able to be used to release radiation. The hand switch provided can be used to release exposures; however, fluoroscopic examinations will not be possible.",
      "recall_initiation_date": "20170928",
      "center_classification_date": "20180223",
      "termination_date": "20200309",
      "report_date": "20180307",
      "code_info": "Serial Numbers: 10006  10010  10017  10025  10044  10050  10051  10055  1044  1078  1081  15000  15001  15018  15021  15032  15044  15047  15056  15058  15060  15076  15088  15091  15105  15114  15116  15131  15144  15148  15149  17002  17004  17007  17011  17032  17050  17061  17062  17074  17078  17084  17142  20001  20002  20005  20009  20015  20024  20030  21004  21006  21017  21022  21028  21033  21039  21047  21052  21054  21075  21087  21089  21095  21096  21112  21114  21117  21132  21135  21150  21151  21173  21175  21178  21187  21188  21200  21201  21205  21208  21216  21225  21273  21274  21278  21281  21283  21294  21308  21309  21317  21327  21331  21337  21345  21347  21351  21352  21353  21363  21370  21377  21379  21384  21393  21400  21407  21414  21415  21421  21423  21425  21430  21440  21442  21444  21447  21448  21449  21450  21461  21462  21467  21473  21475  21484  21488  21494  21498  21500  21508  21509  21515  21519  21520  21523  21524  21536  21543  21550  21551"
    }
  ]
}