      "status": "Ongoing",
      "city": "San Clemente",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "86426",
      "recalling_firm": "ICU Medical, Inc.",
      "address_1": "951 Calle Amanecer",
      "address_2": "N/A",
      "postal_code": "92673-6212",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - U.S. Nationwide distribution including in the states of FL, MD, LA, PA, IN, TX, NY, MO, VA, OH, GA, NC, IL, CA, AZ, AK, NE, DE, WI, CO, UT, ND, MS, MA, MN, WA, SC, OR, KS, MI, NH, NJ, KY, TN, ID, AR, CT, OK, AL, SD, ME, IA, VT, WY, DC, NV, RI, MT, NM, HI, WV and the countries of  Italy, Belgium, United Kingdom, Canada, Germany, Argentina, Australia, Poland, Greece, Switzerland, Israel, France, Finland, Spain, Malaysia, Taiwan, Ireland, Luxembourg, Hong Kong , Brazil, Netherlands, Austria, Sweden, Turkey, Singapore, Slovakia, Croatia, South Africa, Romania, Thailand.",
      "recall_number": "Z-0708-2021",
      "product_description": "76 cm (30\") Appx 3.3 ml, Admin Set, 2 Spiros w/Red Cap, 20 Drop In-Line Drip Chamber w/15 Micron Filter, Bag Hanger, REF: 011-CH3261; 44 cm(17 IN) Appx 2.9 ml,PUR YELLOW TRANS SET,ChemoClave,CK VLV,Spiros,PURPLE CAP, REF: 011-CH3568; Oncology Kit, 5 Gang Stopcock w/5 ChemoClave, Spiros w/Red Cap; 10cm Smallbore Bifuse Ext w/2 ChemoClave, Red Ring, BCV; 297cm Ext w/ChemoClave, REF: 011-CH3600; Oncology Kit, 5 Gang Stopcock w/5 ChemoClave, Spiros w/Red Cap; 10cm Smallbore Bifuse Ext, w/2 ChemoClave, Red Ring, BCV; 597 cm Ext w/ChemoClave, REF: 011-CH3601; Kit, 2 5 Gang Stopcocks w/5 MicroClave Clear, Spiros; 2 Smallbore Bifuse Ext Sets, 2 MicroClave Clear, Red Ring, Check Valve, RMLL; 2 Ext Sets, MicroClave Clear Clipped Together, RMLL, REF: 011-CH3610; Admin Set w/Clave, Spiros, REF: 011-CH3772; Admin Set w/Clave, Spiros, REF: 011-CH3773; Admin Set w/Clave, Spiros, REF: 011-CH3967; 76\" (193 cm) 50 ml, Diluent Set for Channel 2 w/20 Drop Drip Chamber, Spiros, Drop-In Red Cap, REF: 011-CH4011",
      "product_quantity": "4790",
      "reason_for_recall": "Potential for spinning versions of male luers to exhibit leaks due to a molding defect. Fluid leakage may potentially cause delay of infusion, contamination of the fluid path, exposure to hazardous medications, blood loss, exposure to patient blood, or fluid path air-in-line.",
      "recall_initiation_date": "20200831",
      "center_classification_date": "20201223",
      "report_date": "20201230",
      "code_info": "Lots: 4784390, 4784404, 4865931, 4865006, 4590283, 4774210, 4865952, 4711264, 4605516, 4723103, 4882016, 4806569                                                                                                ,"
    }
  ]
}