  "results": [
    {
      "status": "Ongoing",
      "city": "Altstatten Sg",
      "state": "",
      "country": "Switzerland",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "93577",
      "recalling_firm": "Icotec Ag",
      "address_1": "Industriestrasse 12",
      "address_2": "",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "US:  AZ. MI, OH, CA, CO, TX, NY, NC, DE, PA",
      "recall_number": "Z-0700-2024",
      "product_description": "VADER pedicle system, torque wrench, Catalog Number 42-702",
      "product_quantity": "79 devices (US only)",
      "reason_for_recall": "icotec has received notification of one revision that had to be carried out because nut screws had become loose. During the investigation of the device involved, it was detected that the torque limiting function of the wrench used to tighten the nut screws during implantation was not functioning correctly. Due to the malfunction, the nut screws may not have been tightened sufficiently.",
      "recall_initiation_date": "20231128",
      "center_classification_date": "20240110",
      "report_date": "20240117",
      "code_info": "19/01-001, 19/01-013, 19/01-014, 19/01-015, 190830-10, 190830-23, 190830-42, 190830-44, 190830-45, 190830-50, 190830-51, 190830-52, 200131-54, 200131-56, 200131-62, 200131-64, 200131-68, 200131-69, 200131-70, 200131-72, 200131-73, 200131-74, 200221-040R1, 200424-01, 200424-03, 200424-04, 200424-05, 200424-06, 200424-10, 200424-11, 200424-12, 200424-13, 200424-14, 200424-15, 200424-16, 200424-17, 200424-20, 200424-21, 200424-22, 200424-24, 200424-25, 200424-27, 200424-29, 200424-32, 200424-34, 200424-35, 200605-051R1, 200605-052R1, 201002-073R1, 210915-005, 210915-009, 210915-010, 210915-011, 210915-012, 220610-001, 220610-003, 220610-008, 220610-019, 220610-020, 220610-021, 220610-025, 220610-026, 220610-030, 230113-003, 230113-005, 230113-007, 230113-008, 230113-009, 230113-010, 230113-011, 230113-012, 230113-013, 230113-014, 230224-029, 230224-034, 230426-038, 230426-039, 230426-042, 230426-043",
      "more_code_info": ""
    }
  ]
}