  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Center Valley",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "93601",
      "recalling_firm": "Olympus Corporation of the Americas",
      "address_1": "3500 Corporate Pkwy",
      "address_2": "",
      "postal_code": "18034-8229",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide - US Nationwide distribution.",
      "recall_number": "Z-0697-2024",
      "product_description": "Olympus ENDOEYE HD II Rigid Videoscope  Catalog Numbers: WA50040A, WA50042A  -used with a video system center, light source, documentation equipment, monitor, hand instruments, electrosurgical unit, and other ancillary equipment for endoscopy and endoscopic surgery within the thoracic and abdominal cavities including the female reproductive organs.",
      "product_quantity": "WA50040A : 561 units; WA50042A: 857 units",
      "reason_for_recall": "Reports of a pink or green coloration of the image. If the issue is encountered during a procedure, device exchange could potentially result in prolonged surgery; potential visual impacts mucosal injury or bleeding may occur.",
      "recall_initiation_date": "20231106",
      "center_classification_date": "20240110",
      "report_date": "20240117",
      "code_info": "UDI-DI:  WA50040A: 04042761074964 WA50042A: 04042761074971  All Serial Numbers",
      "more_code_info": ""
    }
  ]
}