      "status": "Ongoing",
      "city": "Northfield",
      "state": "IL",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "95821",
      "recalling_firm": "MEDLINE INDUSTRIES, LP - Northfield",
      "address_1": "3 Lakes Dr",
      "address_2": "",
      "postal_code": "60093-2753",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "US and Canada",
      "recall_number": "Z-0696-2025",
      "product_description": "HUDSON RCI mBrace, ET Tube Holder with Bite Block and 2 Point Head Strap, REF DYNJMBRC2B; tracheal tube fixation device",
      "product_quantity": "360 uinits",
      "reason_for_recall": "Complaint defects have been noted regarding the adhesive elements of the device and the frame of the holder putting additional pressure on the patient's cheekbones and upper lip which may cause dermal injury depending on the device positioning.",
      "recall_initiation_date": "20241118",
      "center_classification_date": "20241218",
      "report_date": "20241225",
      "code_info": "UDI/DI 20193489005315 (case), 10193489005318 (each); All Lots",
      "more_code_info": ""
    }
  ]
}