  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Bridgend",
      "state": "N/A",
      "country": "United Kingdom",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "81682",
      "recalling_firm": "ORTHO-CLINICAL DIAGNOSTICS",
      "address_1": "Felindre Meadows",
      "address_2": "N/A",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - US Nationwide domestic distribution. International distribution to Australia, Belgium, Bermuda, Brazil, Canada, Chile, China, Colombia, Denmark, India, France, Japan, Germany, Italy, Mexico, Norway, Poland, Portugal, Russia, Singapore, Spain, Sweden, The Netherlands, and United Kingdom.",
      "recall_number": "Z-0689-2019",
      "product_description": "VITROS NT-proBNP Reagent, Cat No. 6802156    Product Usage:  For in vitro diagnostic use only. For the quantitative measurement of N-terminal pro Brain Natriuretic Peptide (NT-proBNP) in human serum and plasma (EDTA or heparin) using the VITROS ECi/ECiQ Immunodiagnostic Systems, the VITROS 3600 Immunodiagnostic System and the VITROS 5600 Integrated System to aid in the diagnosis of congestive heart failure and for the risk stratification of acute coronary syndrome and congestive heart failure. The test is further indicated as an aid in the assessment of increased risk of cardiovascular events and mortality in patients at risk for heart failure who have stable coronary artery disease. The test can also be used in the assessment of heart failure severity in patients diagnosed with congestive heart failure.",
      "product_quantity": "46598 total, 46062 units in US (100 tests per unit)",
      "reason_for_recall": "The firm is extending their previous July 2018 recall to include additional product codes. Biased results may occur for these additional products at biotin concentrations which are lower than indicated in the current Instructions for Use.",
      "recall_initiation_date": "20181101",
      "center_classification_date": "20190103",
      "report_date": "20190109",
      "code_info": "All expired, in-date, and future lots released."
    }
  ]
}