  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Salt Lake City",
      "state": "UT",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "72923",
      "recalling_firm": "MOOG Medical Devices Group",
      "address_1": "4314 S Zevex Park Ln",
      "address_2": "N/A",
      "postal_code": "84123-7881",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - US (nationwide), Guam, and countries of: Australia, Canada, Croatia, Turkey, New Zealand, Saudi Arabia, Japan, Korea, and Italy.",
      "recall_number": "Z-0688-2016",
      "product_description": "Moog Curlin Infusion, Ambulatory Infusion Pump   The Curlin Ambulatory Infusion System includes Curlin Ambulatory multi-therapy and PainSmart pumps.   Curlin 4000, catalog number 340; Curlin 6000, 6000 CMS, and PainSmart IOD, catalog number 360; and Curlin Pump Calibration Set, catalog number 340-4038.",
      "product_quantity": "11,472",
      "reason_for_recall": "Moog is recalling Curlin 4000, 6000, and PainSmart Infusion Pumps manufactured or serviced by MMDG between March 18, 2015 and November 6, 2015 because the master pumps used to calibrate them were out of tolerance and may cause an over delivery of medications and solutions at a rate of 1.2 - 6.8% faster than programmed.",
      "recall_initiation_date": "20151223",
      "center_classification_date": "20160211",
      "termination_date": "20170221",
      "report_date": "20160217",
      "code_info": "All manufactured or serviced by MMDG between March 18, 2015 and November 6, 2015.  Curlin 4000, catalog number 340; Curlin 6000, 6000 CMS, and PainSmart IOD, catalog number 360; and Curlin Pump Calibration Set, catalog number 340-4038.",
      "more_code_info": ""
    }
  ]
}