  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "San Carlos",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "95574",
      "recalling_firm": "Noah Medical",
      "address_1": "1501 Industrial Rd",
      "address_2": "",
      "postal_code": "94070-4111",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide Distribution: SC, OH, PA, TN, MO, IN, CA, WV, NY, MD, MA, NJ, ID, CT, NC",
      "recall_number": "Z-0679-2025",
      "product_description": "Galaxy bronchoscope (Product GALB-001) used with Instructions for use, Part: 10000752, part of the Noah Medical Galaxy System (GAL-001)",
      "product_quantity": "26",
      "reason_for_recall": "Not all biopsy tools meant to fit bronchoscope channel (2.1mm inner diameter/length greater or equal to 1100mm) will pass through if scope channel positioned in sharp bend (approx. greater than 100-deg.), excessive force may result in tissue injury/pneumothorax; instruments with long rigid distal sections may exhibit insertion resistance when bronchoscope bending section significantly articulated.",
      "recall_initiation_date": "20241025",
      "center_classification_date": "20241212",
      "report_date": "20241218",
      "code_info": "UDI-DI: 00850048825048. IFU Revision: J and all prior versions. All bronchoscope lots used with the following system serial numbers: CRB-1502, CRB-1503, CRB-1508, CRB-1513, CRB-1514, CRB-1515, CRB-1516, CRB-1518, CRB-1519, CRB-1520, CRB-1521, CRB-1522, CRB-1523, CRB-1524, CRB-1525, CRB-1528, CRB-1530, CRB-1532, CRB-1534, CRB-1535, CRB-1537, CRB-1538, CRB1540, CRB-1541, CRB-1543, CRB1554",
      "more_code_info": ""
    }
  ]
}