  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Stockholm",
      "state": "",
      "country": "Sweden",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "81693",
      "recalling_firm": "RAYSEARCH LABORATORIES AB",
      "address_1": "Sveavagen 9",
      "address_2": "",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "AZ  CA  CO  CT  DE  FL  GA  HI  ID  IL  LA  MA  MD  ME  MI  MO  MS  NC  NJ  NY  OH  OR  PA  RI  TN  TX  VA  VT  WA  WI",
      "recall_number": "Z-0679-2019",
      "product_description": "RayStation 3.5, RayStation 4.0, RayStation 4.5, RayStation 4.7, RayStation 5, RayStation 6, RayStation 7 and RayStation 8A",
      "product_quantity": "69",
      "reason_for_recall": "Isocenter shifts when using dose tracking in RayStation 3.5,  RayStation 4.0, RayStation 4.5, RayStation 4.7, RayStation 5, RayStation 6, RayStation 7 and RayStation  8A. In these versions, isocenter shifts in dose tracking are incorrectly interpreted according to the DICOM  patient coordinate system.",
      "recall_initiation_date": "20181112",
      "center_classification_date": "20181222",
      "termination_date": "20200417",
      "report_date": "20190102",
      "code_info": "Software version number:   3.5.0.16 ,  3.5.1.6 ,  4.0.0.14 ,  4.0.1.4 ,  4.0.2.9 ,  4.0.3.4 ,  4.5.0.19 ,  4.5.1.14 ,  4.5.2.7 ,  4.7.0.15 ,  4.7.1.10 ,  4.7.2.5 ,  4.7.3.13 ,  4.7.4.4 ,  4.7.5.4 ,   5.0.0.37 ,  5.0.1.11 ,  5.0.2.35 ,  6.0.0.24 ,  6.1.0.26,  6.1.1.2 ,  6.2.0.7 ,  7.0.0.19 ,  8.0.0.61  or  8.0.1.10",
      "more_code_info": ""
    }
  ]
}