  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Completed",
      "city": "Maple Grove",
      "state": "MN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "86777",
      "recalling_firm": "Boston Scientific Corporation",
      "address_1": "1 Scimed Pl",
      "address_2": "N/A",
      "postal_code": "55311-1565",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationwide",
      "recall_number": "Z-0667-2021",
      "product_description": "IceRod\" 1.5 PLUS Cryoablation Needle",
      "product_quantity": "255 units",
      "reason_for_recall": "Complaint trend regarding needle shaft gas leaks.",
      "recall_initiation_date": "20201118",
      "center_classification_date": "20201222",
      "report_date": "20201230",
      "code_info": "UPN: FPRPR3510    UDIs: 7290104830570, 7290104830570, 7290104830570, 7290104830570, 7290104830570, 7290104830570, 7290104830570, 7290104830570, 7290104830570, 7290104830570, 7290104830570, 7290104830570    Batch Numbers: U1021, U1022, U1014, U1013, U0780, U0781, U0391, U0214, U0008, A2081, A2080, A1531"
    }
  ]
}