  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Sunnyvale",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "72776",
      "recalling_firm": "Intuitive Surgical, Inc.",
      "address_1": "1266 Kifer Rd Bldg 100",
      "address_2": "N/A",
      "postal_code": "94086-5304",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - US (nationwide) and Internationally to Australia, Austria, Belgium, Cyprus, Czech Republic, Denmark, France, Germany, Hong Kong, India, Israel, Italy, Japan, Monaco, Netherlands, Norway, Puerto Rico, Qatar, Romania, Saudi Arabia, Singapore, South Korea, Spain,  Sweden, Switzerland, Taiwan, Turkey, United Kingdom, and Venezuela.    .",
      "recall_number": "Z-0652-2016",
      "product_description": "0 deg, 8mm Endoscope, IS4000:  Model numbers: 470026-34, 470026-40, 470026-41.    General and Plastic Surgery:  The da Vinci Xi endoscope produces high definition (HD), visible light (VIS). The HD video is processed by the system electronics in the Vision Cart and displayed on the Surgeon Console 3D viewer and Vision Cart touchscreen. Endoscopes are available with either a 0 degree or 30 degree tips.",
      "product_quantity": "2624 total all models",
      "reason_for_recall": "This correction to the da Vinci Xi I&A User Manual communicates proper use of the da Vinci Xi 8 mm Endoscopes in combination with the da Vinci 8 mm and 12 mm cannulae, cannula reducers, and monopolar energy instruments.  It informs customers of scenarios that may result in unintended patient harm due to capacitive coupling from monopolar energy sources.",
      "recall_initiation_date": "20151123",
      "center_classification_date": "20160114",
      "termination_date": "20160405",
      "report_date": "20160120",
      "code_info": "Model numbers: 470026-34, 470026-40, 470026-41.  All lots."
    }
  ]
}