{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Waltham",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "72808",
      "recalling_firm": "Perkinelmer",
      "address_1": "940 Winter Street",
      "address_2": "N/A",
      "postal_code": "02451",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - US Nationwide and the countries of   ARGENTINA, AUSTRALIA, AUSTRIA, BELGIUM, BOLIVIA, BRAZIL, CANADA, CHINA, DENMARK, FINLAND, FRANCE, GERMANY, GREAT BRITAIN, GREECE, HONG KONG, INDIA, IRELAND, ISRAEL, ITALY, JAPAN, KOREA, KUWAIT, MYANMAR, NETHERLANDS, NEW ZEALAND, NORWAY, POLAND, PORTUGAL, SAUDI ARABIA, SINGAPORE, SLOVAKIA, SOUTH AFRICA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, and THAILAND.",
      "recall_number": "Z-0637-2016",
      "product_description": "Perkin Elmer WIZARD2 10-detector, 1000 samples.  Product Code: 2470-0200, 3470-0200.  The Wizard gamma counter is intended to detect and count gamma radiation emitted by clinical samples",
      "product_quantity": "22 units",
      "reason_for_recall": "The content of the second #023 barcode ID label in the Barcode ID Label Binder is incorrect.  The second barcode ID label #023 symbol incorrectly identifies as #024 when scanned. If measurement protocols have been programmed for both ID #023 and #024, the protocol associated with barcode ID #024 is executed.  If the error is undetected, the Gamma Counter may produce erroneous results.",
      "recall_initiation_date": "20151208",
      "center_classification_date": "20160112",
      "termination_date": "20170804",
      "report_date": "20160120",
      "code_info": "Serial Numbes:  DG08118243  DG12118732  DG09106871  DG03117554  DG05106373  DG02117481  DG11107153  DG07118106  DG03128981  DG04106316  DG03117574  DG04106296  DG04117674  DG05106414  DG02117455  DG07106607  DG11107048  DG03106163  DG07106604  DG05117805  DG08106759  DG03129008"
    }
  ]
}