{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "San Jose",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "69838",
      "recalling_firm": "Stryker Endoscopy",
      "address_1": "5900 Optical Ct",
      "address_2": "N/A",
      "postal_code": "95138-1400",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - USA (nationwide) and Internationally to Algeria, Argentina, Australia, Bolivia, Brazil, Canada, Chile, China, Colombia, Costa Rica, Ecuador, France, Germany, Greece, Guam, Guatemala, India, Italy, Japan, Jordan, Kuwait, Latin America, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Peru, Philippines, Poland, Portugal, Puerto Rico, Romania, Saudi Arabia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom, and Venezuela.",
      "recall_number": "Z-0624-2015",
      "product_description": "PKG, ROTATING HANDLE, W/RATCHET, INSULATED SHAFT, P/N 0250080234.   Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which allow them to be used for electrosurgery. Laparoscopic surgical manual instruments are designed to be used endoscopically through cannulae to perform cutting, grasping, dissecting, retracting, and manipulating functions. This device does not sustain or support life.",
      "product_quantity": "180,573 units total",
      "reason_for_recall": "The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethylene Oxide cycle sterilization methods",
      "recall_initiation_date": "20141117",
      "center_classification_date": "20141217",
      "termination_date": "20150717",
      "report_date": "20141224",
      "code_info": "42178  42259  42260  42263  42267  42325  42326  42791  70109  713091  72309  90809  1192010  140575  140800  140807  140864  141289  141292  141401  0200154B  0200176B  0200177B  0245386C  0246314C  0246934C  0246939C  0246940C  0246942C  0247044C  0247047C  0247049C  0340029C  0340679C  0340680C  0340681C  0340682C  0340683C  0341237C  0341241C  0341908C  0341912C  4122011  04242012E  04282011E  0441389C  0441455C  0441456C  441457  0441461C  0441464C  0441465C  0441466C  0441467C  0441468C  0441471C  0441472C  0443282C  0443519C  5022011  5112010  5212010  52120101  52120102  52120103  52120104  05232013D  5252010  0540672D  0540867D  0540868D  0540869D  0540871D  0540872D  0540873D  0540874D  0540877D  0540878D  0540880D  0540881D  0540884D  06040590D  06122012E  6162014  6172011  6272012  06272012E  0640590D  0640591D  0640595D  0640596D  0640606D  0640607D  0640608D  640609  0640610D  0640611D  0640612D  0640613D  0640614D  0640615D  0647215D  06CON06064  06CON06123  06CON06666  7022012  7232014  0741767D  0741768D  0741770D  0741772D  0741773D  0741774D  0741776D  0741777D  0741778D  0745271D  0745276D  0745277D  0745278D  0745279D  0745280D  0745281D  0745282D  0840677D  0840768D  0841627D  0841630D  0841636D  0841640D  0841644D  0841648D  0841648E  0841652D  0841690D  0841699D  0841709D  0841720D  0842601D  0846857D  0846858D  0846860D  0846863D  0846866D  0846873D  0846878D  0846883D  0846887D  0846893D  0846900D  0846904D  0846910D  0846919D  0846923D  0846928D  0846935D  0846944D  0846948D  90209  9072010  9232013  9252012  0944346D  0944349D  0944350D  0944355D  0944360D  0944361D  0944365D  0944368D  0944369D  0944370D  0944375D  0944377D  0944380D  0944382D  0944387D  0944391D  0944393D  0944398D  0944402D  0944404D  0944409D  0944790D  0944795D  0944796D  0946948D  1042510D  1044586D  1045207D  1045209D  10456684D  1045680D  1045681D  1045684D  1046169D  1046170D  1046859D  1046861D  1046863D  1046864D  1046868D  1046869D  1046875D  111299  1141741D  1141742D  1142182D  1142183D  1142184D  1143443D  1143444D  1143445D  1145111D  1145112D  1145474D  1145475D  1145637D  1146568D  1146569D  1146570D  1146571D  1240176D  1240177D  1240566D  1240567D  1240956D  1241215D  1241521D  1241522D  1242146D  1242147D  1242164D  1242265  1242265D  1242266  1242266D  1242268D  1242271D  1242655  1242655D  1242656  1242656D  1242657  1242657D  1242658  1243272  1243272D  1243273  1243811  1243812  1244696  1244697  1244697D  1244698  1244699  1245106  1245107  1245107D  1340222  1340222D  1340223  1340223D  1340224  1340225  1340226  1340226D  1340710  1340710D  1341450  1341451  1341452  1341452D  1341453  1341454  1341455  1341455D  1341456  1341850D  1342210  1342211  1342211D  1342212  1342212D  1342214  1342215  1342215D  143465B  1440348  1440349  1440350  1440351  1440352  1441655  151299  180899  2706R  372499  401299  512498"
    }
  ]
}