{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Newark",
      "address_1": "500 Gbc Dr Ms 514",
      "reason_for_recall": "Siemens Healthcare Diagnostics has identified an issue with Dimension Vista software v.3.7 and v.3.8.1 that may result in aliquot well contamination.  If a test is cancelled by the operator while the instrument is aliquotting the sample, the Dimension Vista instrument may continue to deliver the cancelled sample or vial contents to the aliquot well(s). Future samples may be assigned to the same aliquot well(s) and become contaminated with the existing contents or overflow of the aliquot well may occur potentially contaminating a surrounding well(s).  This issue only occurs when an operator manually cancels a test(s) for calibration, quality control (QC) or patient sample while the instrument is in the process of aliquotting the sample.",
      "address_2": "",
      "product_quantity": "1649",
      "code_info": "Dimension Vista software versions 3.7 and 3.8.1",
      "center_classification_date": "20181214",
      "distribution_pattern": "Nationwide/worldwide distribution.",
      "state": "DE",
      "product_description": "Dimension Vista 500, model no. 10284473 an in vitro diagnostic device intended to duplicate manual analytical procedures such as pipetting, mixing, heating, and measuring spectral intensities to determine a variety of analytes in human body fluids.",
      "report_date": "20181226",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Siemens Healthcare Diagnostics, Inc.",
      "recall_number": "Z-0616-2019",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "81478",
      "termination_date": "20200617",
      "more_code_info": "",
      "recall_initiation_date": "20181016",
      "postal_code": "19702-2466",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}