  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Blue Ash",
      "state": "OH",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "93597",
      "recalling_firm": "Megadyne Medical Products, Inc.",
      "address_1": "4545 Creek Rd",
      "address_2": "",
      "postal_code": "45242-2803",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution.  US Nationwide including Puerto Rico; Argentina, Australia, Austria, Bahrain, Belgium, Brazil, Canada, Chile, China, Colombia, Czechia, Denmark, Egypt, Eswatini, Finland, France, Germany, Greece, Hong Kong, Hungary, India, Ireland, Israel, Italy, Japan, Jordan, Kosovo, Kuwait, Lebanon, Malaysia, Martinique, Morocco, Netherlands, New Caledonia, Norway, Philippines, Poland, Portugal, Qatar, R¿union, Romania, Russian , Saudi Arabia, Singapore, Slovakia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, and Vietnam",
      "recall_number": "Z-0598-2024",
      "product_description": "Product Code 0846, MEGADYNE MEGA SOFT Universal Dual Patient Return Electrode, Reusable Patient Return Electrode, Mega Soft Universal Dual Cord",
      "product_quantity": "3,572",
      "reason_for_recall": "Affected product codes are now limited to use in patients age 12 years or older. Megadyne has received reports of patient burn injuries up to and including third-degree burns requiring intervention which may lead to prolonged hospital stay, scarring, and additional surgeries in both pediatric and adult patients.",
      "recall_initiation_date": "20231208",
      "center_classification_date": "20231226",
      "report_date": "20240103",
      "code_info": "UDI-DI 10614559104248  All units within expiry",
      "more_code_info": ""
    }
  ]
}