{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Sparks",
      "state": "MD",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "95636",
      "recalling_firm": "Becton Dickinson & Co.",
      "address_1": "7 Loveton Cir",
      "address_2": "",
      "postal_code": "21152-9212",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide and the countries of Algeria, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Benin, Bhutan, Bolivia, Bosnia and Herzegovina, Botswana, Brazil, Brunei, Bulgaria, Burkina Faso, Burundi, Cambodia, Cameroon, Canada, Chile, China, Colombia, Costa Rica, Cote d'Ivoire, Croatia, Cyprus, Czech Republic, Democratic Republic of the Congo, Denmark, DOM, Dominica, Dominican Republic, Ecuador, Egypt, El Salvador, Estonia, Ethiopia, Faroe Islands, Fiji, Finland, France, French Guiana, Gambia, Georgia, Germany, Ghana, Greece, Guadeloupe, Guam, Guatemala, Guinea, Guyana, Haiti, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Jamaica Japan, Jordan, Kazakhstan, Kenya, Kosovo, Kuwait, Kyrgyzstan, Laos, Latvia, Lebanon Liberia, Libyan Arab Jamahiriya, Liechtenstein, Lithuania, Luxembourg, Macau, Madagascar Malawi, Malaysia, Maldives, Mali, Malta, Martinique, Mauritania, Mauritius, Mayotte Mexico, MKD, Moldova, Mongolia, Morocco, Mozambique, Namibia, Nepal, Netherlands Antilles, New Caledonia, New Zealand, Nicaragua, Nigeria, Northern Mariana Islands, Norway, Oman, Pakistan, Panama, Papua new Guinea, Paraguay, Peru, Philippines PNG, Poland, Portugal, Puerto Rico, Qatar, Reunion, Romania, Russia, Rwanda, Saint Helena San Marino, Saudi Arabia, Senegal, Serbia, Seychelles, Sierra Leone, Singapore, Slovakia Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sudan, Swaziland, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Tajikistan, Tanzania, United Republic of Thailand, Timor-Leste, Tonga, Trinidad And Tobago, Tunisia, Turkey, Turkmenistan Uganda, Ukraine, United Arab Emirates, United Kingdom, Uruguay Uzbekistan, Vietnam, Yemen, Zambia, Zimbabwe.",
      "recall_number": "Z-0596-2025",
      "product_description": "BACTEC 9240 System  Version or Model: 445475  Catalog Number: 445475",
      "product_quantity": "584 units (313 US, 271 OUS)",
      "reason_for_recall": "BD confirmed that product service credentials used by some BD technical support teams to access certain BD products were accessed by an unauthorized actor. Until these product service credentials are updated, there is a risk of unauthorized access that may impact the confidentiality, integrity and/or availability of the relevant products and associated data.",
      "recall_initiation_date": "20241031",
      "center_classification_date": "20241203",
      "report_date": "20241211",
      "code_info": "Catalog No: 445475; UDI-DI: 00382904454753; Serial Numbers: DB5631, DB 6970, DB6857, DB6859, DB6122, DB3043, MD1710, DB4372, DB6043, DB2249, DB5030, DB2882, DB3090, MD2046, DB4927, DB5825, DB2828, DB2835, DB3251, DB3245, DB6893, DB3145, DB3211, DB5876, DB5906, DB4284, DB5948, DB3143, DB3204, DB6891, PM4385R, DB5761, DB4811, DB4718, DB3231, DB2786, DB4793, DB2322, DB6522, DB4820, DB2216, DB6558, DB4113, DB5168, DB2544, DB6674, DB4263, DB4842, DB6369, DB6900, DB4828, DB4418, DB5204, DB5282, DB4286, DB5569, DB5626, FS1472, FS1473, FS1230, FS1234, DB2702, DB4367, DB6655, DB6920, DB4365, DB2219, FT1004, DB6156, DB6961, DB2421, DB5244, DB2530, DB2844, DB4890, DB5146, DB5443, DB6011, DB5940, DB6628, DB3318, DB2417, MD1748, DB6315, DB4134, FS1320, DB3307, DB4304, MD2027, DB4152, MD1851, FS1280, DB2152, DB6243, DB4680, FS1004, FS1094, FS1001, DB6570, DB6391, DB6393, DB3004, DB4706, DB4248, DB6392, DB4260, DB6573, DB6571, DB2840, DB6584, DB2974, DB6634, DB4267, DB4440, DB2548, DB4449, DB6491, DB6903, DB4892, DB5529, DB2925, DB6595, DB5538, DB4271, DB6936, DB4987, DB2337, DB3389, DB3340, DF1540, DB2332, DB7013, DB4492, FS1178, FS1519, DB4565, DB4589, DB4028, DB6430, DB3076, DB6258, MD1730, MD1810, DB4586, DB5340, DB6499, DB5347A, DB5398, DB6603, MD2086, DB6113, DB2884, DB4733, DB4739, DB2804, DB4631, DB2276, DB4506, DB3024, DB6123, DB6590, DB4557, DB6281, 94707, DB6426, DB6657, DB6933, DB6658, DB4936, DB6324, DB2561, DB5865, DB5163, DB5005, DB6745, DB5200, DB5878, DB6709, DB4869, DB6058, DB6646, DB6645, DB6843, DB5094, DB4917, DB6845, DB6863, DB7074, DB7003, DB7071, DB7004, DB6994, DB7041, DB7133, DB7132, DB7167, DB6755, DB7187, DB7144, DB7142, DB6451, DB6689, DB7199, DB7182, DB7203, DB7136, DB5880, DB6896, DB6464, DB6679, FS1426, MD1634, DB1679, DB7050, DB6342, DB6872, DB2657, DB6576, DB6719, DB6938, DB4377, DB7000, DB6733, FS1400, DB2556, MD1750, DB6237, DB5602, DB7007, DB7008, DB6883, DF1563, DB6950, DB6559, DB6951, DF1562, DB6918, DB5858, DB5350, DB6041, DB4998, DB6484, DB6505, DB5363, DB6792, DB7212, DB5381, DB5422, DB5682, DB5653, DB5429, DB5784, FS1494, DB2516, DB4662, DB4665, MD1793, MD1794, DB4763, DB6809, MD1916, DB5905, MD1610, MD1840, DB2160, DB7178, DB6302, DB6165, MD1925, MD2056, MD2122, DB5596, DB5919, DB5861, DB5862, DB5813, DB5937, DB5089, DB6034, DB2403, MD1619, DB7157, DB2453, DB2424, DB5215, DB1059, DB2238, DB4020, DB5740, DB5065, DB6858, DB5955, DB2244, DB5021, DB3086, DB4776, 97R4YC2, DB6296, DB5328, DB2245, DB2483, DB5329, DB2946, DB2945, DB5491, DB4348, DB5319, DB4299, DB4484, MD2033, DB3062, DB3279, DB3349, DB4149, DB 5844, DB 5845, DB5847, DB 5997, DB 5846, DB5888, DB5683, DB4805, DB6802, DF1547, DB6218, DB6685, RC1164, DB2212, DB3212, DB4868, DB4891, DB4478, DB4742, DB5096, DB4855, DB6262, DB5668, DB2384, DB6024, DB6071, DB7101, DB7098, DB7100, DB6018, DB6036, DB2939, DB2967, DB5286, DB2146, FS1073, DB7120, DB5522, DB2977, DB6096, DB6483, DB6604, DB6640, DB6750, DB6670, DB6814, DB6815, DB7099, DB6990, DB7122, DB7110, DB6736, DB7139, DB7140, DB7190, DB7180, DB7191, DB7194, DB7216, DB7204, DB7205, DB7197, DB7173, DB2312, DB6597, DB5077, DB3291, DB4538, DB6155, DB6485, DB6486, DB6487, DB6488, DB7062, DB7185, DB4344, DB4453, DB4952, MD1771, MD1877, DB4138, DB6142, DB6477, MD1847, FS1470, FS1476, DB5084, FS1244, MD2047, DB5302, DB7030, DB2170, DB4816, DB5620, DB5169, DB5507, DB5737, DB5202, DB6811, DB7034, DB2220, DB5525, DB2221, DB6715, DB6708, DB6734, DB6781, DB2588, DB6780, DB7054, DB6268, DB2204, DB5696, DB6403, DB2737, DB5643, DB6717, DB2660, DB6591, DB6911, DB2633, DB5743, DB6852, DB2357, DB2758, DF1552, DB6790, TBD1234, DB6786, DB6323, DB6996, DB6941, DB6850, DB6907, DB6980, DB6952, DB5449, DB5375, DB5373, DB7076, DB4234, DB5608, DB5407, DB5652, DB5430, DB5512, DB4550, DB5594, DB7097, DB6213, DB5550, MD1832, DB2493, DB5083, MD1917, DB5734, DB7175, DB5911, DB5756, DB7176, DB7179, DB5800, DB5914, MD1611, DB2705, DB2377, MD1615, DB2378, DB6032, DB6054, DB6861, DB6494, DB4055, DB2845, DB6357, DB4083, DB4225, DB6121, DB4635, DB4078, DB6383, DB6671, DB6381, DB2819, DB3154, FS1011, DB2966, FS1193, DB6532, DB2818, DB6291, DB6283, MD1791, DB6617, DB6457, DB3124, DB4835, DB6169, DB4682, DB4596, DB6212, DB4649, DB4366, DB7096, DB6929, DB6978, DB6913, DB6319, FS1485, FS1341, DB2178, DB2619, FS1270, DB6240, MD1690, DB6317, DB6239, DB6300, DB4872, DB4441, DB2998, DB6581, DB6540, DB4451, DB6635, DB6699, DB2335, FS1461, DB6160, FS1204, DB5325, DB6161, FS1520, DB4206, DB4024, DB4765, FS1419, DF1596, MD1624, DB4619, DB6242, DB6496, DB4321, DB6331, DB6241, DB6667, DB4759, DB4322, DB4326, DB4752, DB4338, DB6339, DB3054, DB6989, DB4594, DB4938, DB4939, DB7022, DB7023, DB5177, DB5232, DB5205, DB5242, DB5274, DB7040, DB6905, DB2163, DB6065, DB6993, DB2620,",
      "more_code_info": ""
    }
  ]
}