{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "San Jose",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "69838",
      "recalling_firm": "Stryker Endoscopy",
      "address_1": "5900 Optical Ct",
      "address_2": "N/A",
      "postal_code": "95138-1400",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - USA (nationwide) and Internationally to Algeria, Argentina, Australia, Bolivia, Brazil, Canada, Chile, China, Colombia, Costa Rica, Ecuador, France, Germany, Greece, Guam, Guatemala, India, Italy, Japan, Jordan, Kuwait, Latin America, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Peru, Philippines, Poland, Portugal, Puerto Rico, Romania, Saudi Arabia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom, and Venezuela.",
      "recall_number": "Z-0596-2015",
      "product_description": "PKG, BIPOLAR FORCEPS, FENESTRATED, 33CM, P/N 0250080123  Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which allow them to be used for electrosurgery. Laparoscopic surgical manual instruments are designed to be used endoscopically through cannulae to perform cutting, grasping, dissecting, retracting, and manipulating functions. This device does not sustain or support life.",
      "product_quantity": "180,573 units total",
      "reason_for_recall": "The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethylene Oxide cycle sterilization methods",
      "recall_initiation_date": "20141117",
      "center_classification_date": "20141217",
      "termination_date": "20150717",
      "report_date": "20141224",
      "code_info": "06CON06113  0816243/0817410  0913221/0817442  0913222/0817442  1015164/1014813  1015164/1015778  1016405/1016061  1016405/1016400  1016405/1016401  1113335/1113339  1213126/1114543  1213127/1114543  1213127/1213123  1214001/1313047  1313947/1313946  0817389/0817444  0913565/0913326  0915791/0915061  0915791/1013608  0915894/0915061  0915895/0915061  1014635/1015778  1015165/1013609  1015165/1015778  1015773/1015778  1015774/1015778  1015862/1016061  1016124/1016061  1016125/1016062  1016128/1016061  1016406/1016400  1016408/1016400  1016408/1016826  1016562/1016400  1113333/1016063  1113333/1113018  1113333/1113136  1113333/1113339  1115969/1115651  1116112/1115651  1116113/1115651  1116114/1115651  1116115/1115651  1116256/1115651  1213128/1213731  1213973/1213978  1213974/1213731  1213974/1213978"
    }
  ]
}