{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Watertown",
      "state": "SD",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "97861",
      "recalling_firm": "Pivotal Health Solutions, Inc.",
      "address_1": "3003 9th Ave Sw",
      "address_2": "N/A",
      "postal_code": "57201-8014",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Distribution to US nationwide, Canada and Japan.",
      "recall_number": "Z-0588-2026",
      "product_description": "Quantum Intersegmental Table, Model/Catalog Number: AMQ400H; heated massage table",
      "product_quantity": "464 units",
      "reason_for_recall": "The heat function may become damaged over time due to patients pressing on the areas with heat function components. Damage to components can lead to overheating which can lead to smoldering and melting of the vinyl cushion.",
      "recall_initiation_date": "20251009",
      "center_classification_date": "20251124",
      "report_date": "20251203",
      "code_info": "All serial numbers.",
      "more_code_info": ""
    }
  ]
}