  "results": [
    {
      "status": "Terminated",
      "city": "Bloomington",
      "state": "IN",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "84212",
      "recalling_firm": "Cook Inc.",
      "address_1": "750 N Daniels Way",
      "address_2": "N/A",
      "postal_code": "47404-9120",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Distribution to US states of  AL, CA, CT, FL, GA, IL, MA, MD, MI, MO, NC, NJ, NY, OH, OK, PA, SC, TN, TX, and VA, and Brazil.",
      "recall_number": "Z-0588-2020",
      "product_description": "CrossCath Support Catheter, RPN CXC3.4-2.2-18-90-P-NS-0, Order Number G51575   For diagnostic and interventional vascular procedures.",
      "product_quantity": "6",
      "reason_for_recall": "A manufacturing error may cause the radiopaque marker bands to be too loose on certain CXC3.0 CrossCath Support Catheters (compatible with 0.014 wire guides) and too tight on certain CXC3.4 CrossCath Support Catheters (compatible with 0.018 wire guides).",
      "recall_initiation_date": "20191108",
      "center_classification_date": "20191209",
      "termination_date": "20210201",
      "report_date": "20191218",
      "code_info": "Lot Number 9945303  UDI: (01)00827002515758(17)221308(10)9945303",
      "more_code_info": ""
    }
  ]
}