  "results": [
    {
      "status": "Ongoing",
      "city": "Watertown",
      "state": "SD",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "97861",
      "recalling_firm": "Pivotal Health Solutions, Inc.",
      "address_1": "3003 9th Ave Sw",
      "address_2": "N/A",
      "postal_code": "57201-8014",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Distribution to US nationwide, Canada and Japan.",
      "recall_number": "Z-0587-2026",
      "product_description": "Quantum Intersegmental Table, Model/Catalog Number: Q400; heated massage table",
      "product_quantity": "34 units",
      "reason_for_recall": "The heat function may become damaged over time due to patients pressing on the areas with heat function components. Damage to components can lead to overheating which can lead to smoldering and melting of the vinyl cushion.",
      "recall_initiation_date": "20251009",
      "center_classification_date": "20251124",
      "report_date": "20251203",
      "code_info": "All serial numbers.",
      "more_code_info": ""
    }
  ]
}