{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "West Chester",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "86713",
      "recalling_firm": "Synthes (USA) Products LLC",
      "address_1": "1301 Goshen Pkwy",
      "address_2": "N/A",
      "postal_code": "19380-5986",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationwide  Foreign:countries: United Arab Emirates, Austria, Australia, Belgium, Switzerland, China, Czech Republic, Germany, Spain, United Kingdom, Ireland, Israel, Italy, Singapore,",
      "recall_number": "Z-0586-2021",
      "product_description": "DePuy Synthes 17.5MM REAMER HEAD FOR RIA 2 STERILE-Orthopedic manual surgical instrument  Part Code: 03.404.031S",
      "product_quantity": "515 US; 150 OUS",
      "reason_for_recall": "RIA 2 reamer head breakage exceeds the expected occurrence rate resulting in adverse tissue reaction, soft tissue damage, and/or bone damage, when the recommended surgical techniques are not followed",
      "recall_initiation_date": "20201119",
      "center_classification_date": "20201215",
      "termination_date": "20220701",
      "report_date": "20201223",
      "code_info": "All Lots  UDI: 10886982274076"
    }
  ]
}