  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "San Jose",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "69818",
      "recalling_firm": "Boston Scientific Corp",
      "address_1": "150 Baytech Dr",
      "address_2": "N/A",
      "postal_code": "95134-2302",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution -- US, including the states of MA, AL, NC, and OK; and, the countries of Germany, France, and Great Britain.",
      "recall_number": "Z-0580-2015",
      "product_description": "IntellaMap Orion High Resolution Mapping Catheter;  Material number: M004RC64S0; Catalog number RC64S;    Is indicated for electrophysiological mapping (recording or stimulating only) of the cardiac structures of the heart.",
      "product_quantity": "184 units",
      "reason_for_recall": "Some IntellaMap Orion High Resolution Mapping Catheters are exhibiting a failure mode in which a glue bond on the Deployment Interlock Button has been observed to fail in the field.",
      "recall_initiation_date": "20141120",
      "center_classification_date": "20141210",
      "termination_date": "20160811",
      "report_date": "20141217",
      "code_info": "Lot numbers: 17075550, 17075981, 17083913, 17088826, 17094081, 17101668, 17208445, 17218467, 17251661, 17258617, 17274937, 17292792, and 17308586. Expiry Dates from June 25, 2015 to September 24, 2015."
    }
  ]
}