  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Grafton",
      "state": "WI",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "95746",
      "recalling_firm": "Gauthier Biomedical, Inc.",
      "address_1": "2221 Washington St",
      "address_2": "",
      "postal_code": "53024-9506",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide distribution in the state of MA.",
      "recall_number": "Z-0572-2025",
      "product_description": "TriALTIS TORQUE LIMITER, RATCHET, 11.3 Nm,  Model Number 5161-00-5048 (G919509)",
      "product_quantity": "160 units",
      "reason_for_recall": "An out of specification component may result in the device exhibiting torque output higher than the design specification.",
      "recall_initiation_date": "20241111",
      "center_classification_date": "20241127",
      "report_date": "20241204",
      "code_info": "Model No 5161-00-5048; UDI-DI 00858325005688; Lot Numbers: GB153869 and GB156110",
      "more_code_info": ""
    }
  ]
}