  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Carlsbad",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "86692",
      "recalling_firm": "Ra Medical Systems, Inc.",
      "address_1": "2070 Las Palmas Dr",
      "address_2": "N/A",
      "postal_code": "92011-1518",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Distribution to states of: AL, AZ, CA, FL, IL, IN, KY, LA, MD, NC, NV, NY, OK, OR, TX, and WA; and to OUS country: Japan.",
      "recall_number": "Z-0569-2021",
      "product_description": "DABRA Laser (RA-308 Excimer Laser)    Intended Use: For use in ablating a channel in occlusive peripheral vascular disease. Indications: For crossing chronic total occlusions in patients with symptomatic infrainguinal lower extremity vascular disease.",
      "product_quantity": "36 devices",
      "reason_for_recall": "Due to footswitch not meeting required specification for protection from ingress of solids or liquids.  If spilling of liquids near or on the footswitch, it may fail in the 'on' position and allow the laser to continue to lase.",
      "recall_initiation_date": "20201016",
      "center_classification_date": "20201214",
      "termination_date": "20230111",
      "report_date": "20201223",
      "code_info": "Model Number RA-308  Product/Catalog Number: 1200-0001-01  Serial Numbers:  RA00027, RA00046, RA00049, RA00050, RA00054, RA00056, RA00057, RA00058, RA00064, RA00069, RA00073, RA00081, RA00083, RA00084, RA00085, RA00089, RA00097, RA00098, RA00114, RA00115, RA00120, RA00121, RA00125, RA00127, RA00135, RA00138, RA00148, RA00153, RA00154, RA00160, RA00164, RA00165, RA00167, RA00170, RA00175, and RA00177."
    }
  ]
}