  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Redmond",
      "state": "WA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "93234",
      "recalling_firm": "Physio-Control, Inc.",
      "address_1": "11811 Willows Rd Ne",
      "address_2": "N/A",
      "postal_code": "98052-2003",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY.",
      "recall_number": "Z-0567-2024",
      "product_description": "AC Power Adapter for use with the LIFEPAK 15 Monitor/Defibrillator, REF: 11140-000098",
      "product_quantity": "18,732",
      "reason_for_recall": "AC Power Adapter (ACPA) may not charge the monitor/defibrillator batteries and may not allow the monitor/defibrillator to power on while plugged into the ACPA.",
      "recall_initiation_date": "20231106",
      "center_classification_date": "20231215",
      "report_date": "20231227",
      "code_info": "UDI-DI: 00883873940204, Adapters manufactured between 11/3/2021 - 4/24/2023"
    }
  ]
}