  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Santa Ana",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "86681",
      "recalling_firm": "Medtronic Heart Valves Division",
      "address_1": "1851 E Deere Ave",
      "address_2": "N/A",
      "postal_code": "92705-5720",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - U.S. Nationwide distribution and the countries of Panama, Denmark, Austria, Germany, Finland, Sweden, Colombia, India, Mexico, Brazil, Spain, United Kingdom, Australia, Ireland, Italy, Argentina, Japan, Poland, Russian Federation, Hungary, Kazakhstan, Luxembourg, Portugal, Norway, Egypt, Saudi Arabia.",
      "recall_number": "Z-0560-2021",
      "product_description": "Evolut (TM) PRO System.  With the following model numbers:      a. EVOLUTPRO-23, Labeled as: VLV EVOLUTPRO-23 PRO COMM 29L OUS, VLV EVOLUTPRO-23 BLUE 29L COMM OUS, VLV EVOLUTPRO-23 BLUE COMM JPN, VLV EVOLUTPRO-23 BLUE 6L COMM OUS, VLV EVOLUTPRO-23 PRO COMM 29L CAN, VLV EVOLUTPRO-23 PRO COMM 29L OUS SA, VLV EVOLUTPRO-23 PRO COMM 29L AUS;      b. EVOLUTPRO-23-US, Labeled as: VLV EVOLUTPRO-23-US BLUE COMM US, VLV EVOLUTPRO-23-US PRO BLNK US;      c. EVOLUTPRO-26, Labeled as: VLV EVOLUTPRO-26 PRO COMM 29L OUS, VLV EVOLUTPRO-26 BLUE 6L COMM OUS, VLV EVOLUTPRO-26 PRO COMM 29L AUS, VLV EVOLUTPRO-26 BLUE COMM JPN, VLV EVOLUTPRO-26 BLUE 29L COMM OUS, VLV EVOLUTPRO-26 PRO COMM 29L CAN, VLV EVOLUTPRO-26 PRO COMM 29L OUS SA;      d. EVOLUTPRO-26-US, Labeled as: VLV EVOLUTPRO-26-US BLUE COMM US, VLV EVOLUTPRO-26-US PRO BLNK US;      e. EVOLUTPRO-29, Labeled as: VLV EVOLUTPRO-29 PRO COMM 29L OUS, VLV EVOLUTPRO-29 BLUE 29L COMM OUS, VLV EVOLUTPRO-29 BLUE 6L COMM OUS, VLV EVOLUTPRO-29 PRO COMM 29L AUS, VLV EVOLUTPRO-29 BLUE COMM JPN, VLV EVOLUTPRO-29 PRO COMM 29L CAN, VLV EVOLUTPRO-29 PRO COMM 29L OUS SA;      f. EVOLUTPRO-29-US, Labeled as: VLV EVOLUTPRO-29-US BLUE COMM US, VLV EVOLUTPRO-29-US PRO BLNK US      g. , Labeled as:",
      "product_quantity": "71,030 units",
      "reason_for_recall": "Potential for Transcatheter Aortic Valve (TAV) leaflet damage when performing a post-implant balloon dilatation (PID).",
      "recall_initiation_date": "20201023",
      "center_classification_date": "20201210",
      "termination_date": "20221122",
      "report_date": "20201216",
      "code_info": "All GTINs and associated serial numbers (refer to Attachment A-Product List as of 22-Oct-2020)."
    }
  ]
}