  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Best",
      "state": "N/A",
      "country": "Netherlands",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "93416",
      "recalling_firm": "PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.",
      "address_1": "Veenpluis 4-6",
      "address_2": "N/A",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide - US Nationwide and the countries of Albania, Algeria, Argentina, Australia,  Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, CANADA, Cayman Islands, Chile, China, Colombia, Costa Rica, Croatia, Cuba, Curacao, Cyprus, Czech Republic, Denmark, Dominican Republic,  Ecuador, Egypt, Estonia, Fiji, Finland, France, French Polynesia, Gabon, Georgia,  Germany, Ghana, Greece, Guadeloupe, Guatemala, Guinea, Honduras, Hong Kong,  Hungary, Iceland, India, Indonesia, Iran, IRAQ, Ireland, Israel, Italy, Japan, Jordan,  Kazakhstan, Kenya, Korea, Republic of Kosovo, Kuwait, Latvia, Latvia, Republic of Lebanon, Libya, Lithuania, Macao, Macedonia, Malaysia, Maldives, Martinique, Mauritius,  Mexico, Mongolia, Morocco, Myanmar, Nepal, Netherlands, New Caledonia, New Zealand,  Nicaragua, Norway, Oman, Pakistan, Palestine, State of, Panama, Papua New Guinea,  Paraguay, Peru, Philippines, Poland, Portugal, Puerto Rico, Qatar, R¿union, Romania,  Russian Federation, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sudan, Sweden, Switzerland, Syrian Arab Republic, Taiwan,  Thailand, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uruguay,  Uzbekistan, Venezuela, Vietnam, Yemen.",
      "recall_number": "Z-0558-2024",
      "product_description": "Philips Allura  systems with monoplane fixed ceiling mounts:  722001 Allura Xper FD10C  722003 Allura Xper FD10  722006 Allura Xper FD20  722010 Allura Xper FD10  722012 Allura Xper FD20  722015 Allura Xper FD20 OR Table  722016 Integris H5000 C / Allura 9C  722018 Integris Allura 9  722022 Allura Xper FD10 OR Table  722023 Allura Xper FD20 OR Table  722026 Allura Xper FD10  722028 Allura Xper FD20  722033 Allura Xper FD10 OR Table  722035 Allura Xper FD20 OR Table  722043 Integris Allura 15 & 12 (monoplane)  72246 Poly G - OMCP - VISUB - CCD (H5000)",
      "product_quantity": "9991 units",
      "reason_for_recall": "Ceiling mounted L-arm contains a rotation cover that may potentially be susceptible to falling if a collision between the L-arm and other hospital equipment (i.e., an operating light) were to occur result in injury or potential sterility issues due to the cover becoming loose",
      "recall_initiation_date": "20231103",
      "center_classification_date": "20231215",
      "report_date": "20231227",
      "code_info": "Model/Name/UDI-DI: 722001 Allura Xper FD10C;  722003 Allura Xper FD10;  722006 Allura Xper FD20 (01)00884838059115(21); 722010 Allura Xper FD10 (01)00884838059030(21); 722012 Allura Xper FD20 (01)00884838059054(21); 722015 Allura Xper FD20 OR Table;  722016 Integris H5000 C / Allura 9C ;  722018 Integris Allura 9; -  722022 Allura Xper FD10 OR Table;  722023 Allura Xper FD20 OR Table (01)00884838059085(21;) 722026 Allura Xper FD10 (01)00884838054189(21); 722028 Allura Xper FD20 (01)00884838054202(21); 722035 Allura Xper FD20 OR Table (01)00884838054240(21); 722043 Integris Allura 15 & 12 (monoplane)."
    }
  ]
}