  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Mansfield",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "89418",
      "recalling_firm": "Covidien Llc",
      "address_1": "15 Hampshire St",
      "address_2": "N/A",
      "postal_code": "02048-1113",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide and the countries of Germany, Israel, Italy, Portugal, Spain, Chile, Singapore, and China.",
      "recall_number": "Z-0557-2022",
      "product_description": "Chameleon PTA Balloon Dilation Catheter 8 mm x 40 mm, Chameleon PTA Balloon Catheter with Injection Port 8 mm x 40 mm; Cardiovascular.  Product Numbers CH08-40-75GB and CH08-40-75US",
      "product_quantity": "1,994 devices",
      "reason_for_recall": "The firm has received six (6) complaint reports from customers indicating difficulty removing the balloon protector sleeve from the catheter (e.g., excessive force required to remove the protector sleeve). If the balloon is damaged while removing the protector  sleeve and the catheter is subsequently used in a patient procedure, patient harm may potentially occur, including perforation of vessels, foreign body in patient, embolism, risks  associated with surgical intervention, or delay to treatment/therapy.",
      "recall_initiation_date": "20211217",
      "center_classification_date": "20220203",
      "report_date": "20220209",
      "code_info": "UDI: 10884521812253, 07290016745047, 10884521812291  For CFN CH08-40-75US: GTIN                    Batch 07290016745047 21904374 07290016745047 21904432 07290016745047 22000673 07290016745047 22000709 07290016745047 22000817 07290016745047 22000880 07290016745047 22001030 07290016745047 22001061 07290016745047 22001191 07290016745047 22001310 07290016745047 22001417 07290016745047 22001448 07290016745047 22001728 07290016745047 22002086 07290016745047 22002088 10884521812253 22002498 07290016745047 22002583 10884521812253 22002603 07290016745047 22003037 07290016745047 22100003 07290016745047 22100321 07290016745047 22100322 07290016745047 22101089 07290016745047 22101090 10884521812291 22102189"
    }
  ]
}