  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Kyoto",
      "state": "",
      "country": "Japan",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "95632",
      "recalling_firm": "Shimadzu Corp. Analytical & Measuring Instruments Div.",
      "address_1": "1 Chome 1",
      "address_2": "Nishinokyokuwabara-Cho;  Nakagyo; Hon-Machi; Higashiyama",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide - US Nationwide distribution in the states of MS, AR, TX, CA, MA, MI, FL, OK, WI, IN, MN, OH, NJ, NY, IL, KY, NV, HI, SD, ID, MI, UT, WA, MO, VA, GA and the countries of DE, CN, JP, UY.",
      "recall_number": "Z-0553-2025",
      "product_description": "Autosampler, Models: SIL-20ACXR CL, SIL-20ACHT CL, SIL-30AC CL",
      "product_quantity": "179",
      "reason_for_recall": "There have been cases of failure of the main board due to condensation water drained from the Peltier dehumidifier, leaked out from the tray, resulting in a short circuit. In the serious case, the main board and adjacent cables are partly charred. In the minor case, the Y motor and its surrounding die cast part become rusty. These failures could result in delay in patient sample testing.",
      "recall_initiation_date": "20241009",
      "center_classification_date": "20241126",
      "report_date": "20241204",
      "code_info": "Model/UDI-DI/Serial Number: SIL-20ACXR CL/04540217420131/L20995250001, L20995250002, L20995250003, L20995250004, L20995250005, L20995250006, L20995250007, L20995250008, L20995350010, L20995350011, L20995350012, L20995350013, L20995350014, L20995350015, L20995450016, L20995450017, L20995450018, L20995450019, L20995450020, L20995450021, L20995450022, L20995450023, L20995450024, L20995450025, L20995450026, L20995550027, L20995550028, L20995550029, L20995550030, L20995650031, L20995650032, L20995650033, L20995650034, L20995650035, L20995650036, L20995650037, L20995650038, L20995650039, L20995650040, L20995650041, L20995650042, L20995650043, L20995650044, L20995650045, L20995650046, L20995560001, L20995660002, L20995660003, L20995660004, L20995660005, L20995660006, L20995660007, L20995660008, L20995660009, L20995660010, L20995660011, L20995660012, L20995660013, L20995660014, L20995660015, L20995660016, L20995660017, L20995660018, L20995660019, L20995660020, L20995660021, L20995660022, L20995660023, L20995660024, L20995660025, L20995660026, L20995660027, L20995660028, L20995660029, L20995660030, L20995660031, L20995660032, L20995660033, L20995660034, L20995660035, L20995660036, L20995660037, L20995660038, L20995660039, L20995660040, L20995660041, L20995660042, L20995660043, L20995660044, L20995660045, L20995660046, L20995660047, L20995660048, L20995660049, L20995660050, L20995660051, L20995660052, L20995660053, L20995660054, L20995660055, L20995660056, L20995660057, L20995250009, L20995270001, L20995270002, L20995270003, L20995270004, L20995270005, L20995370006, L20995370007, L20995370008, L20995370009, L20995370010, L20995370011, L20995370012, L20995370013, L20995370014, L20995370015, L20995370016, L20995370017, L20995370018, L20995370019, L20995370020, L20995370021, L20995370022, L20995370023, L20995370024, L20995370025, L20995470028, L20995770032.  SIL-20ACHT CL/04540217420124/L20985350001, L20985450002, L20985750003, L20985750004.  SIL-30AC CL/04540217420155/L21005250001, L21005250002, L21005250003, L21005250004, L21005250005, L21005250006, L21005250007, L21005350008, L21005350009, L21005350010, L21005350011, L21005350012, L21005350013, L21005350014, L21005350015, L21005350016, L21005350017, L21005350018, L21005350019, L21005350020, L21005350021, L21005350022, L21005350023, L21005350024, L21005450025, L21005550026, L21005550027, L21005650028, L21005650029, L21005750030, L21005750031, L21005570001, L21005570002, L21005670003, L21005670004, L21005670005, L21005670006, L21005670007, L21005670008, L21005770009, L21005770010, L21005770011, L21005770012, L21005770013, L21005770014.",
      "more_code_info": ""
    }
  ]
}