      "status": "Ongoing",
      "city": "Sylmar",
      "state": "CA",
      "country": "United States",
      "classification": "Class III",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "95557",
      "recalling_firm": "Abbott Medical",
      "address_1": "15900 Valley View Ct",
      "address_2": "",
      "postal_code": "91342-3577",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Visit",
      "distribution_pattern": "Worldwide - U.S. Nationwide distribution in the states of ALABAMA, ALASKA, ARKANSAS, CALIFORNIA, COLORADO, CONNECTICUT, DISTRICT OF COLUMBIA, FLORIDA, GEORGIA, ILLINOIS, INDIANA, KENTUCKY, MICHIGAN, MISSISSIPPI, MISSOURI, MONTANA, NEBRASKA, NEW JERSEY, NEW YORK, NORTH CAROLINA, OHIO, OREGON, PENNSYLVANIA, SOUTH CAROLINA, TENNESSEE, TEXAS, VIRGINIA, WASHINGTON and WISCONSIN. The country of Australia.",
      "recall_number": "Z-0552-2025",
      "product_description": "Merlin@home with the MerlinOnDemand capability Transmitter REF EX1150    Merlin@home with the MerlinOnDemand\" (MoD) capability is a transmitter  configured for multiple patient use. It can receive information from an  implantable heart device in unpaired mode. With MoD capability, the transmitter  will retrieve full follow-up data or retrieve data for display on the Merlin.net\"  Patient Care Network (PCN). The MoD capability enables clinicians to do the  following:  . support remote care patients in emergency rooms and heart failure clinics  . triage patients for in clinic follow-ups  . support patients in a clinical setting with Merlin.net PCN",
      "product_quantity": "Pending",
      "reason_for_recall": "Due to heart transmitter not being able to upgrade to the current software version in the field after experiencing a power interruption during a prior over-the-wire software download.",
      "recall_initiation_date": "20241001",
      "center_classification_date": "20241126",
      "report_date": "20241204",
      "code_info": "Model Number: EX1150 UDI-DI code: Pending Serial Numbers: Pending",
      "more_code_info": ""
    }
  ]
}