  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Tustin",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "86864",
      "recalling_firm": "Canon Medical System, USA, INC.",
      "address_1": "2441 Michelle Dr",
      "address_2": "N/A",
      "postal_code": "92780-7047",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "FAX",
      "distribution_pattern": "US Nationwide distribution. OUS: Unknown",
      "recall_number": "Z-0549-2021",
      "product_description": "Canon Aquilion Prime SP, Multislice Helical CT Scanner, TSX-303B/8L - Product Usage: intended to acquire and display cross-sectional volumes of the whole body, including the head.",
      "product_quantity": "60 units",
      "reason_for_recall": "A software problem has been identified  which could result in the CT Scanner not proceeding to the next actual scan even though automatic start of the next scan is specified. This could result in the CT scan failing  resulting in rescanning and reinjection of contrast medium.",
      "recall_initiation_date": "20201109",
      "center_classification_date": "20201209",
      "termination_date": "20221216",
      "report_date": "20201216",
      "code_info": "Serial Numbers: 8LA2052063,         8LA19Z2034 ,         8LA19X2004,       8LA19Z2027,          8LA19Z2029,          8LA2052062,         8LA2042055,         8LA19Z2032,          8OA2052059,        8LA2042051,         8LA2032044,         8LA19X2002,         1LB1922175,         8LA2042050,         8LA19X2007,         8LA2012037,         8LA19Y2021,          8LA19Y2014,          8LA2042056,         8LA19Z2023,          8LA2052067,         8LA19Z2031,          8LA19Y2019,          8LA19Z2030,          8OA19Y2013,         8LA19Z2028,          8LA19Y2012,         8LA19Y2020,         8LA2032042,         8LA2052066,         8LA19X2003,         8LA19Y2022,         8LA19X2008,         8LA2042047,         8LA19Z2024,          8LA2032046,         8LA19Z2033,          8OA2042053,        8LA2012036,        8LA2042054,         8LA19Z2025,          8LA19Z2026,          8LA2022041,         8LA19X2010,         8LA2032043,         8OA19X2001,        8OA2052061,        8OA2042049,        8LA19Y2016,        8LA2022039,         8LA19Y2015,         8LA2012035,  8LA2032045,      8LA19Y2018,      8LA19X2005,      8LA19X2006,      8LA19X2009,      8LA19Y2011,      8LA19Y2017,      8LA2012038,      8LA2052060,      8LA2052058 ,     8LA2052064,      8LA2052065,      8LA2052068,      8LA2022040,      8OA2042052,     8LA2042048,      8LA2052057"
    }
  ]
}