  "results": [
    {
      "status": "Terminated",
      "city": "Cleveland",
      "state": "OH",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "81569",
      "recalling_firm": "Philips Medical Systems (Cleveland) Inc",
      "address_1": "595 Miner Rd",
      "address_2": "N/A",
      "postal_code": "44143-2131",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide Distribution",
      "recall_number": "Z-0546-2019",
      "product_description": "Brilliance 16 Slice (Air), Model 728246    Product Usage:  Computed Tomography X-ray systems intended to produce cross-sectional Images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes. These devices may include signal analysts and display equipment, patient and equipment supports, components, and accessories",
      "product_quantity": "6,343 units in total",
      "reason_for_recall": "There is a potential for separation of fiberglass resin which may cause the gantry front cover and the metal support on the cover to fail/fall",
      "recall_initiation_date": "20181018",
      "center_classification_date": "20181130",
      "termination_date": "20210803",
      "report_date": "20181212",
      "code_info": "System Serial Number:  646  1554  3007  3010  3011  3012  3013  3014  3015  3016  3017  3018  3020  3023  3027  3029  3030  3031  3038  3040  3041  3043  3047  3050  3057  3059  3070  3072  3082  3085  3090  3095  3103  3110  3111  3117  3123  3127  3135  3141  3149  3150  3160  3161  3162  3163  3164  3170  3179  3185  3189  3191  3202  3205  3216  3217  3219  3224  3227  3244  3247  3279  3286  3290  3291  3293  3294  3295  3300  3302  3304  3306  3314  3324  3329  3331  3335  3356  3359  3373  3376  3382  3392  3393  3407  3421  3423  3435  3441  3442  3444  3458  3465  3466  3473  3478  3482  3490  3514  3521  3526  3535  3563  3565  3570  3588  3603  3616  3624  3647  3650  3665  3673  3720  3721  3769  3841  4004  4013  4015  4017  4018  4019  4024  4025  4036  4037  4038  4041  4042  4045  4046  4051  4055  5001  5002  5004  5005  5006  5007  5008  5010  5011  5012  5013  5014  5015  5016  5017  5018  5019  5020  5021  5022  5023  5024  5025  5026  5027  5028  5029  5030  5031  5032  5033  5034  5035  5036  5037  5038  5040  5041  5042  5044  5045  5046  5049  5050  5051  5052  5053  5054  5055  5058  5059  5060  5061  5062  5064  5065  5066  5067  5068  5069  5070  5072  5073  5074  5075  5076  5077  5078  5079  5081  5082  5083  5084  5085  5086  5087  5088  5089  5090  5091  5092  5093  5094  5095  5097  5098  5099  5101  5102  5103  5104  5105  5106  5107  5108  5109  5110  5111  5112  5113  5114  5115  5116  5117  5118  5119  5120  5121  5122  5123  5124  5125  5126  5127  5128  5129  5130  5131  5132  5133  5134  5135  5136  5137  5138  5139  5140  5141  5142  5143  5145  5146  5147  5148  5149  5150  5151  5152  5153  5155  5158  5159  5160  5161  5162  5163  5164  5165  5166  5167  5168  5169  5170  5171  5171  5172  5173  5174  5175  5176  5177  5178  5179  5180  5181  5183  5184  5185  5186  5187  5188  5189  5190  5191  5192  5193  5194  5195  5196  5197  5198  5199  5200  5201  5202  5203  5204  5205  5206  5206  5207  5208  5209  5210  5211  5212  5213  5214  5215  5216  5217  5218  5219  5220  5221  5222  5223  5224  5225  5226  5227  5228  5229  5230  5231  5232  5233  5234  5235  5236  5237  5238  5239  5240  5241  5242  5243  5245  5246  5247  5248  5249  5250  5252  5253  5254  5255  5256  5257  5258  5259  5260  5261  5262  5263  5264  5265  5266  5267  5268  5269  5270  5271  5272  5273  5274  5275  5276  5278  5279  5281  5282  5283  5285  5286  5287  5288  5289  5290  5291  5292  5293  5294  5295  5296  5297  5298  5299  5300  5301  5302  5303  5304  5305  5306  5307  5308  5309  5311  5312  5313  5314  5315  5316  5317  5318  5319  5320  5321  5322  5323  5324  5325  5326  5327  5328  5329  5330  5331  5332  5333  5334  5335  5336  5337  5338  5339  5340  5341  5342  5343  5344  5346  5347  5348  5349  5350  5351  5352  5353  5354  5355  5357  5358  5359  5360  5361  5362  5363  5364  5365  5366  5367  5368  5369  5370  5371  5372  5373  5375  5376  5377  5378  5379  5380  5381  5382  5384  5386  5387  5388  5389  5391  5393  5394  5396  5397  5398  5399  5400  5401  5402  5403  5404  5405  5406  5407  5408  5409  5410  5411  5412  5414  5415  5416  5417  5418  5419  5420  5421  5422  5423  5424  5425  5426  5427  5428  5429  5431  5432  5433  5434  5435  5437  5438  5439  5440  5441  5442  5443  5444  5445  5446  5447  5448  5449  5450  5451  5452  5453  5454  5456  5457  5458  5459  5460  5461  5462  5463  5464  5465  5466  5468  5469  5470  5471  5472  5473  5474  5475  5476  5477  5478  5479  5480  5481  5482  5483  5484  5485  5486  5487  5488  5489  5490  5491  5492  5493  5494  5496  5497  5498  5501  5502  5503  5504  5505  5506  5507  5508  5509  5510  5511  5513  5514  5515  5516  5517  5518  5519  5520  5521  5522  5523  5524  5525  5526  5528  5529  5530  5531  5532  5533  5534  5535  5536  5537  5538  5539  5540  5541  5543  5544  5545  5546  5547  5548  5549  5551  5552  5553  5554  5555  5556  5556  5557  5558  5559  5560  5561  5562  5563  5564  5565  5566  5567  5568  5569  5570  5571  5572  5573  5574  5575  5576  5577  5578  5579  5580  5581  5582  5583  5584  5585  5586  5587  5588  5589  5590  5591  5592  5593  5594  5595  5596  5597  5598  5599  5600  5601  5602  5603  5604  5605  5606  5607  5608  5609  5610  5611  5612  5613  5614  5615  5616  5617  5618  5619  5620  5621  5622  5623  5624  5625  5626  5627  5628  5629  5630  5631  5632  5633  5634  5635  5636  5637  5638  5639  5640  5641  5642  5643  5644  5645  5646  5647  5648  5649  5650  5651  5652  5653  5654  5655  5656  5657  5658  5659  5660  5661  5662  5663  5664  5665  5666  5667  5668  5669  5670  5671  5672  5673  5673  5674  5675  5676  5677  5678  5679  5680  5681  5682  5683  5684  5685  5686  5687  5688  5689  5690  5691  5692  5693  5694  5695  5696  5697  5698  5699  5700  5701  5702  5703  5704  5705  5706  5707  5708  5709  5710  5711  5712  5713  5714  5715  5716  5717  5718  5719  5720  5721  5722  5723  5724  5725  5726  5727  5728  5729  5730  5731  5732  5733  5734  5735  5736  5737  5738  5739  5740  5741  5742  5743  5744  5745  5746  5747  5748  5749  5750  5751  5752  5753  5754  5755  5756  5757  5758  5759  5760  5761  5762  5763  5764  5765  5766  5767  5768  5769  5770  5771  5772  5773  5774  5775  5776  5777  5778  5779  5780  5781  5782  5783  5784  5785  5786  5787  5788  5789  5790  5791  5792  5793  5794  5795  5796  5797  5798  5799  5800  5801  5802  5803  5804  5805  5806  5807  5808  5809  5810  5811  5812  5813  5814  5815  5816  5817  5818  5819  5820  5821  5822  5823  5824  5825  5826  5827  5828  5829  5830  5831  5832  5833  5834  5835  5836  5837  5838  5839  5840  5841  5842  5843  5844  5845  5846  5847  5848  5849  5850  5851  5852  5853  5854  5855  5856  5857  5858  5859  5860  5861  5862  5863  5864  5865  5866  5867  5868  5869  5870  5871  5872  5873  5874  5875  5876  5877  5878  5879  5880  5881  5882  5883  5884  5885  5886  5887  5888  5889  5891  5892  5893  5894  5895  5896  5897  5898  5899  5900  5901  5902  5903  5904  5905  5906  5907  5908  5909  5910  5911  5912  5913  5914  5915  5916  5917  5918  5919  5920  5921  5922  5923  5924  5925  5926  5927  5928  5929  5930  5931  5932  5933  5934  5935  5936  5937  5938  5939  5940  5941  5942  5943  5944  5945  5946  5947  5948  5949  5950  5951  5952  5953  5954  5955  5956  5957  5958  5959  5960  5961  5962  5963  5964  5965  5966  5967  5968  5969  5970  5971  5972  5973  5974  5975  5976  5977  5978  5979  5980  5981  5982  5983  5984  5985  5986  5987  5988  5989  5990  5991  5992  5993  5994  5995  5996  5997  5998  5999  6000  6001  6002  6003  6004  6005  6006  6007  6008  6010  6011  6012  6013  6014  6015  6016  6017  6018  6020  6021  6022  6023  6024  6025  6026  6027  6028  6029  6030  6031  6032  6033  6034  6035  6036  6037  6038  6039  6040  6041  6042  6043  6044  6045  6046  6047  6048  6049  6050  6051  6052  6053  6054  6055  6056  6057  6058  6059  6060  6061  6062  6063  6064  6065  6066  6067  6068  6069  6070  6071  6072  6073  6074  6075  6076  6077  6078  6079  6080  6081  6082  6083  6084  6085  6086  6087  6088  6089  6090  6091  6092  6093  6094  6095  6096  6097  6098  6101  6102  6104  6105  6106  6107  6108  6109  6110  6111  6112  6113  6114  6115  6116  6117  6118  6119  6120  6121  6122  6123  6124  6125  6126  6127  6128  6129  6130  6131  6132  6133  6134  6135  6136  6137  6138  6139  6141  6142  6143  6144  6145  6146  6147  6148  6149  6150  6150  6151  6152  6153  6155  6156  6157  6158  6159  6160  6161  6162  6163  6164  6165  6166  6167  6168  6169  6170  6171  6172  6173  6174  6175  6176  6177  6178  6179  6180  6181  6182  6183  6184  6185  6186  6187  6188  6189  6190  6191  6192  6193  6194  6195  6196  6197  6198  6199  6200  6201  6202  6203  6204  6205  6206  6207  6208  6209  6210  6211  6212  6213  6214  6215  6216  6217  6218  6219  6220  6221  6222  6223  6224  6225  6226  6227  6228  6229  6230  6231  6232  6233  6234  6235  6236  6237  6238  6239  6240  6241  6242  6243  6244  6245  6246  6247  6248  6249  6250  6251  6252  6253  6254  6255  6256  6257  6258  6259  6260  6261  6262  6263  6264  6265  6266  6267  6268  6269  6270  6271  6272  6273  6274  6275  6276  6277  6278  6279  6280  6281  6282  6283  6284  6285  6286  6287  6288  6289  6290  6291  6292  6293  6294  6295  6296  6297  6298  6299  6300  6301  6302  6303  6304  6305  6306  6307  6308  6309  6310  6311  6312  6313  6314  6315  6316  6317  6318  6319  6320  6321  6322  6323  6324  6325  6326  6327  6328  6329  6330  6331  6332  6333  6334  6335  6336  6337  6338  6339  6340  6341  6342  6343  6344  6345  6346  6347  6348  6349  6350  6351  6352  6353  6354  6355  6356  6357  6358  6359  6360  6361  6362  6363  6364  6365  6366  6367  6368  6369  6370  6371  6372  6373  6374  6375  6376  6377  6378  6379  6380  6382  6383  6384  6385  6386  6387  6388  6389  6392  6393  6394  6395  6397  6398  6399  6400  6401  6402  6403  6404  6405  6406  6407  6408  6409  6410  6411  6412  6413  6414  6415  6416  6417  6418  6419  6420  6421  6422  6423  6424  6425  6426  6427  6428  6429  6430  6431  6432  6433  6434  6435  6436  6437  6438  6439  6440  6441  6442  6443  6444  6445  6446  6447  6448  6449  6450  6451  6452  6453  6454  6455  6456  6457  6458  6459  6460  6461  6462  6463  6464  6465  6466  6467  6468  6469  6470  6471  6472  6473  6474  6475  6476  6477  6478  6479  6480  6481  6482  6483  6484  6485  6486  6487  6488  6489  6490  6491  6492  6493  6494  6495  6496  6497  6498  6499  6500  6501  6502  6503  6504  6505  6506  6507  6508  6509  6510  6511  6512  6513  6514  6515  6516  6517  6518  6519  6520  6521  6522  6523  6524  6525  6526  6527  6528  6529  6530  6531  6532  6533  6534  6535  6536  6537  6538  6539  6540  6541  6542  6543  6544  6545  6546  6547  6548  6549  6550  6551  6552  6553  6554  6555  6556  6557  6558  6559  6560  6561  6562  6563  6564  6565  6566  6567  6568  6569  6570  6571  6572  6573  6574  6575  6576  6577  6578  6579  6580  6581  6582  6583  6584  6585  6586  6587  6588  6589  6590  6591  6592  6593  6594  6595  6596  6597  6598  6599  6600  6601  6602  6603  6604  6605  6606  6607  6608  6609  6610  6611  6612  6613  6614  6615  6616  6617  6618  6619  6620  6621  6622  6623  6624  6625  6626  6627  6628  6629  6630  6631  6632  6633  6634  6635  6636  6637  6638  6639  6640  6641  6642  6643  6644  6645  6646  6647  6648  6649  6650  6651  6652  6653  6654  6655  6656  6657  6658  6659  6660  6661  6662  6663  6664  6665  6666  6667  6668  6669  6670  6671  6672  6673  6674  6675  6676  6677  6678  6679  6680  6681  6682  6683  6684  6685  6686  6687  6688  6689  6690  6691  6692  6693  6694  6695  6696  6697  6698  6699  6700  6701  6702  6703  6704  6705  6706  6707  6708  6709  6710  6711  6712  6713  6714  6715  6716  6717  6718  6719  6720  6721  6722  6723  6724  6725  6726  6727  6728  6729  6730  6731  6732  6733  6734  6735  6736  6737  6738  6740  6741  6742  6743  6744  6745  6746  6747  6748  6749  6750  6751  6752  6753  6754  6755  6756  6757  6758  6759  6760  6761  6762  6763  6764  6765  6766  6767  6768  6769  6770  6771  6772  6773  6774  6775  6776  6777  6778  6779  6780  6781  6782  6783  6784  8731  9788  13456  30009  30016  30019  30020  30023  30028  30033  30041  30044  30051  30060  30061  30063  30067  30069  30071  30076  30081  30090  30091  30093  30097  30101  30103  30109  30115  30119  30122  32111  32366  40001  40002  40003  40008  40013  40017  40018  40020  40022  40025  40027  40030  40032  40033  40035  50001  50002  50003  50004  50005  50006  50007  50008  50009  50010  50011  50012  50013  50014  50015  50016  50017  50018  50019  50021  50022  50023  50024  50025  50026  50027  50028  50029  50030  50031  50032  50033  50034  50035  50036  50037  50038  50040  50041  50042  50043  50044  50045  50046  50047  50048  50049  50050  50051  50052  50053  50054  50055  50056  50057  50058  50059  50060  50061  50062  50063  50064  50065  50066  50067  50068  50069  50070  50071  50072  50073  50074  50075  50076  50077  50078  50079  50080  50081  50082  50083  50084  50085  50086  50087  50088  50089  50090  50091  50092  50093  50094  50095  50096  50097  50098  50099  50100  50101  50102  50103  50104  50105  50106  50107  50108  50109  50110  50111  50112  50113  50114  50115  50116  50117  50118  50119  50120  50121  50122  50123  50124  50125  50126  50127  50128  50129  50131  50132  50133  50135  50136  50137  50138  50139  50140  50141  50142  50143  50144  50145  50146  50147  50148  50149  50150  50151  50152  50153  50154  50155  50156  50157  50158  50159  50160  50161  50162  50163  50164  50165  50166  50167  50168  50169  50170  50171  50172  50173  50175  50176  50177  50178  50179  50180  50181  50182  50183  50184  50185  50186  50187  50188  50189  50190  50191  50192  50193  50194  50195  50196  50197  50198  50199  50200  50201  50202  50204  50205  50206  50207  50208  50209  50210  50211  50212  50213  50214  50215  50216  50217  50218  50219  50220  50221  50222  50223  50224  50225  50226  50227  50228  50229  50230  50231  50232  60001  60005  60007  60013  60015  60020  60024  60028  80055  80064  260001  260002  260004  260005  260006  260007  260008  260009  260010  260011  260012  260013  260014  260015  260016  260017  260018  260019  260020  260021  260022  260023  260024  260025  260026  260027  260028  260029  260030  260031  260032  260033  260034  260035  260036  260037  260038  260039  260040  260041  260042  260043  260044  260045  260046  260047  260048  260049  260050  260051  260052  260053  260054  260055  260056  260057  260058  260059  260060  260061  260062  260063  260064  260065  260066  260067  260068  260069  260070  260071  260072  260073  260074  260075  260076  260077  260078  260079  260080  260081  260082  260083  260084  260085  260086  260087  260088  260089  260090  270001  500006  553849  557516  628001  6785259  39252638  44129584  51792079  52245700  52245739  52245778  52314824  55139347  55957657  55976120  55976122  57761585  58051570  58685532  59681375  62984841  63147122  69140169  69580027  70521287  70521324  70570226  71978589  11787-01  260038_  260039_  4043*  4702_08  50023_  50057A  50134_  5087-1  5094_DEL  5238A  5389A  5666-OLD  5916-OLD  6140A  6140-OLD  6244/13897/79922  6264_  6762_  CT5781/EBW11689  ESPREVENTIVOSCT  IBA_5104  IBA_5134  IBA_5162  SN50187  SN5682  SN5836  SN6620  STOR_DEMO_43574835  SZ50219  SZ50220  XXX_5870"
    }
  ]
}