  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Cambridge",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "86848",
      "recalling_firm": "Philips North America Llc",
      "address_1": "222 Jacobs St",
      "address_2": "N/A",
      "postal_code": "02141-2289",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide distribution in the state of SD and the countries of China, England, Germany, Hong Kong, Portugal.",
      "recall_number": "Z-0544-2021",
      "product_description": "Philips IntelliVue X2 Multi measurement Server: M3002A (865039) - Product Usage: intended for monitoring and recording of, and to generate alarms for, multiple physiological parameters of adults, pediatrics, and neonates in a hospital environment and during patient transport inside and outside of the hospital environment.",
      "product_quantity": "16 total: 2 units US ; 14 OUS",
      "reason_for_recall": "Speakers on IntelliVue X2 and IntelliVue MP2 may fail,  identified by the display of a  Speaker Malfunct.  INOP or by the absence of audible sound, may result in delayed treatment of the patient.",
      "recall_initiation_date": "20201118",
      "center_classification_date": "20201208",
      "termination_date": "20210722",
      "report_date": "20201216",
      "code_info": "Serial Numbers:  US:  DE73607502  DE73607498    OUS:   System Serial Number  DE95038935  DE73608191  DE73608209  DE73601727                       DE73604240  DE73601547  DE73601596  DE73604182  DE83617535  DE73608259  DE83617500  DE83617522  DE73601651"
    }
  ]
}