  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Rochester",
      "state": "NY",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "93366",
      "recalling_firm": "Ortho-Clinical Diagnostics, Inc.",
      "address_1": "100 Indigo Creek Dr",
      "address_2": "N/A",
      "postal_code": "14626-5101",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide. Global Distribution.",
      "recall_number": "Z-0543-2024",
      "product_description": "VITROS 3600 Immunodiagnostic System Product Code 6802783 and Refurbished Product Code 6802914 Running Software Version 3.8.0",
      "product_quantity": "63 units",
      "reason_for_recall": "During planned monitoring of data post-release of Software Version 3.8.0 an increased occurrence of software issues relating to unresponsive subsystems, unresponsive user interface (sometimes referred to as a screen freeze), and/or printer disconnections was identified. The consequence of these issues would be a delay in reporting results due to the required system shutdowns to recover the system. This delay in reporting results could occur during any assay, including which the speed of the result is deemed critical or STAT and could result in a delay of treatment in a patient. Due to the release strategy of this software version, the user could not update the system but a field engineer had to manually install each affected system.",
      "recall_initiation_date": "20231010",
      "center_classification_date": "20231213",
      "report_date": "20231220",
      "code_info": "Product Code: 6802783,  UDI-DI: 10758750002979;  Product Code: 6802914,  UDI-DI: 10758750007103;  Serial Numbers: J36000109 J36000294 J36001012 J36001489 J36000292 J36000340 J36001024 J36001495 J36000484 J36000348 J36001027 J36001513 J36000489 J36000446 J36001051 J36001515 J36000955 J36000511 J36001088 J36001519 J36001151 J36000517 J36001108 J36001528 J36001167 J36000580 J36001111 J36001535 J36001222 J36000688 J36001112 J36001570 J36001283 J36000753 J36001117 J36001573 J36001330 J36000861 J36001237 J36001575 J36001348 J36000877 J36001245 J36001579  J36000920 J36001246 J36001581  J36000923 J36001401 J36001582  J36000929 J36001403 J36001583  J36000933 J36001427 J36001600  J36000941 J36001451 J36001602  J36000942 J36001488 J36001618  J36001000"
    }
  ]
}