{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Deerfield",
      "state": "IL",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "89017",
      "recalling_firm": "Baxter Healthcare Corporation",
      "address_1": "1 Baxter Pkwy",
      "address_2": "N/A",
      "postal_code": "60015-4625",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution.  US nationwide Puerto Rico and St. Thomas.  There was government/military distribution.  Countries of Canada, Bahamas, Barbados, Bermuda, Guyana, Jamaica, and Trinidad and Tobago.",
      "recall_number": "Z-0529-2022",
      "product_description": "Baxter Spectrum IQ Infusion Pumps, Product code 3570009.",
      "product_quantity": "241,304 units",
      "reason_for_recall": "There is the potential for reduced or non-delivery of medication, in some cases without alerting the user via pump alarm. This may occur as a result of incorrect administration set setup and/or incomplete resolution of upstream occlusion alarms when using Spectrum V8 and Spectrum IQ infusion pumps.",
      "recall_initiation_date": "20211229",
      "center_classification_date": "20220128",
      "report_date": "20220209",
      "code_info": "All serial numbers, GTIN 00085412610900."
    }
  ]
}