      "status": "Terminated",
      "city": "Southbridge",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "66862",
      "recalling_firm": "Karl Storz Endoscopy America Inc",
      "address_1": "15 Wells St",
      "address_2": "N/A",
      "postal_code": "01550-4503",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationwide Distribution.",
      "recall_number": "Z-0527-2014",
      "product_description": "Broncho Flexible Fiberscope, models: 11001BN1, 11002BD1, 11004BC1, and 11009BC1.    \" Examination and treatment of the tracheobronchial tree Contraindications for Bronchoscopy \" Uncontrolled angina, uncontrolled heart failure, or serious uncontrolled ventricular arrhythmias \" Uncontrolled hypertension \" Bleeding disorder or use of anticoagulants.",
      "product_quantity": "141 units",
      "reason_for_recall": "KARL STORZ Endoscopy America, Inc. is recalling the Flexible Broncho Fiberscope or a Flexible Intubation Fiberscope with a suction port because the material used in the suction channel portion was thermoplastic polyurethane (TPU), which is not listed in STERRAD systems claims for interior endoscope lumens.",
      "recall_initiation_date": "20131113",
      "center_classification_date": "20131217",
      "termination_date": "20140327",
      "report_date": "20131225",
      "code_info": "C1140, C1140E, C1160, C1160E, QKC1750, QKC1750E, QKC1694, QKC1694E"
    }
  ]
}