  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Northfield",
      "state": "IL",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "97846",
      "recalling_firm": "Medline Industries, LP",
      "address_1": "3 Lakes Dr",
      "address_2": "",
      "postal_code": "60093-2753",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "US Nationwide distribution.",
      "recall_number": "Z-0522-2026",
      "product_description": "MEDLINE convenience kits labeled as:    1) BASIC NEURO PACK, REF CDS780202R;   2) ACDF CDS, REF CDS941061M;   3) NEURO PACK-LF, REF DYNJ0038440C;   4) NEURO VP SHUNT/VAGAL NERVE-RF, REF DYNJ21925S;   5) NEURO-ORTHO EXTRAS, REF DYNJ44882M;   6) LAMINECTOMY, REF DYNJ903870F.",
      "product_quantity": "1315 kits",
      "reason_for_recall": "Medline Industries, LP is issuing a recall for specific item(s) and lot(s) of Medline Kits containing SafeAir Smoke Evacuation Pencils. Stryker Instruments has expanded the scope of their recall of SafeAir Smoke Evacuation Pencils previously initiated in October 2024, adding an additional 47 lots to the scope of the recall. The initial recall was due to the potential for product to activate without manual input when plugged into a power source or remain active after buttons are released.",
      "recall_initiation_date": "20250930",
      "center_classification_date": "20251114",
      "report_date": "20251126",
      "code_info": "1) REF CDS780202R, UDI/DI 10195327418618 (each) 40195327418619 (case), Lot Numbers:  23LMG063;   2) REF CDS941061M, UDI/DI 10193489445589 (each) 40193489445580 (case), Lot Numbers: 24ABG356, 24CBH117, 24DBJ347;    3) REF DYNJ0038440C, UDI/DI 10195327356699 (each) 40195327356690 (case), Lot Numbers: 23JMC956, 23KMF585, 24AMB974;    4) REF DYNJ21925S, UDI/DI 10195327105020 (each) 40195327105021 (case), Lot Numbers: 23EBS176, 23GBF899, 23HBZ085;    5) REF DYNJ44882M, UDI/DI 10195327382803 (each) 40195327382804 (case), Lot Numbers: 23FDA522, 23GDB903;    6) REF DYNJ903870F, UDI/DI 10195327239268 (each) 40195327239269 (case), Lot Numbers:  23IBD913, 23IBD914, 23JBL889, 24ABE596, 24BBL042, 24CBH196, 24DBU825.",
      "more_code_info": ""
    }
  ]
}