  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Malvern",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "89260",
      "recalling_firm": "Siemens Medical Solutions USA, Inc",
      "address_1": "40 Liberty Blvd",
      "address_2": "",
      "postal_code": "19355-1418",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide distribution.",
      "recall_number": "Z-0517-2022",
      "product_description": "Artis icono floor with software VE20C-diagnostic imaging angiography system  Model: 11327700",
      "product_quantity": "23 units",
      "reason_for_recall": "Four potential software issues with Artis pheno and Artis icono systems with software VE20C: 1.\tUpdated calibration data not saved with measurement after scene+/-; 2.\tNo x-ray possible, system shutdown/restart might be required during intervention; 3. Corrupted Image during Roadmap;  4.\tUnintended shutdown of Imaging System with UPS (Uninterruptable Power Supply) option",
      "recall_initiation_date": "20211012",
      "center_classification_date": "20220121",
      "termination_date": "20240411",
      "report_date": "20220202",
      "code_info": "Serial Numbers: 170319 170026 170343 170305 170307 170331 170309 170310 170023 170304 170328 170334 170336 170025 170346 170051 170333 170332 170312 170027 170303 170322 170053  UDI: 04056869149325",
      "more_code_info": ""
    }
  ]
}