  "results": [
    {
      "country": "United States",
      "city": "Miami",
      "address_1": "11800 SW 147th Ave",
      "reason_for_recall": "When the UniCel DxH Slidemaker Stainer and UniCel DxH Slidemaker Stainer II dispense probe is slightly bent, there is a potential for cell carryover from the previous sample, and erroneous results could be reported from the slide review.  Erroneous results may be reported from the laboratory which may trigger unnecessary or inappropriate patient treatment.",
      "address_2": "",
      "product_quantity": "1,056 (418 US, and 638 OUS)",
      "code_info": "Catalog Number 775222.  Serial numbers: AS39017, AS40018, AT35013, AT35014, AT35015, AT35017, AT35018, AT40020, AT40021, AT40022, AT40024, AT40025, AT40027, AT40030, AT45031, AT45032, AT45034, AT45035, AT49037, AT49041, AT49044, AT49045, AT49046, AT49047, AT49048, AT49049, AT49050, AT49051, AT49055, AT49057, AT49058, AT49059, AT49060, AT49061, AT49062, AT49063, AT49064, AT49065, AU01004, AU01005, AU01007, AU01009, AU01010, AU01011, AU01012, AU01014, AU01015, AU01016, AU01017, AU01020, AU01021, AU01022, AU01023, AU01024, AU01025, AU01028, AU01029, AU01030, AU08041, AU08043, AU08044, AU08047, AU08048, AU08049, AU08050, AU11051, AU11053, AU11057, AU11058, AU11059, AU11061, AU11062, AU11063, AU11064, AU11066, AU11068, AU11072, AU13073, AU13075, AU13077, AU13079, AU14083, AU14084, AU14085, AU14086, AU14087, AU14088, AU14090, AU16104, AU16111, AU16112, AU19113, AU19115, AU19118, AU19119, AU19120, AU20123, AU20124, AU20125, AU20128, AU20129, AU20130, AU20131, AU21133, AU21136, AU21137, AU21141, AU23146, AU23147, AU23149, AU23150, AU23151, AU23152, AU28153, AU28154, AU28160, AU28161, AU29163, AU29165, AU29167, AU29168, AU29169, AU29172, AU30173, AU30178, AU30180, AU30181, AU30182, AU35183, AU35185, AU35186, AU35187, AU35188, AU36189, AU36192, AU36193, AU36196, AU36198, AU36200, AU36201, AU36202, AU36203, AU36204, AU40205, AU40207, AU40209, AU40210, AU40211, AU40213, AU40214, AU40215, AU40216, AU40217, AU40218, AU40221, AU43223, AU43225, AU43226, AU43228, AU43229, AU43231, AU43232, AU43233, AU43234, AU43235, AU43237, AU43240, AU49254, AU49256, AU49259, AV01003, AV01004, AV01005, AV01007, AV01009, AV01010, AV01011, AV01012, AV01013, AV01014, AV01016, AV06018, AV06020, AV06023, AV06026, AV06028, AV06030, AV10033, AV10034, AV10035, AV10036, AV10038, AV10039, AV10042, AV10044, AV10045, AV12047, AV12049, AV12054, AV12055, AV12056, AV14057, AV14059, AV14061, AV14062, AV14063, AV16072, AV16073, AV16075, AV16077, AV16078, AV16080, AV16082, AV16083, AV16086, AV16088, AV16089, AV16090, AV16091, AV21092, AV21093, AV21095, AV21096, AV22097, AV22100, AV22102, AV22105, AV22106, AV23107, AV23109, AV23110, AV23111, AV23112, AV23113, AV23116, AV27117, AV27118, AV27120, AV27121, AV36127, AV36128, AV36129, AV36130, AV39131, AV39132, AV39133, AV39134, AV39135, AV40136, AV40137, AV40139, AV40140, AV40143, AV40145, AV42147, AV42149, AV42150, AV42151, AV42152, AV42153, AV42154, AV42155, AV44156, AV44157, AV44158, AV44160, AV44161, AV44162, AV44163, AV44164, AV47167, AV47168, AV47169, AV47170, AV48171, AV48173, AV48174, AV48176, AV48177, AV48178, AV48179, AV49181, AV49183, AV49184, AV49185, AV49187, AV49188, AV49189, AV51191, AV51192, AW01001, AW01002, AW01007, AW01008, AW01010, AW01011, AW01012, AW01013, AW01014, AW01015, AW06016, AW06018, AW06019, AW06020, AW06021, AW06023, AW06024, AW06025, AW06026, AW06027, AW06028, AW06029, AW09031, AW09032, AW09033, AW09034, AW09035, AW10036, AW10037, AW10038, AW10039, AW10040, AW10041, AW10044, AW10045, AW14046, AW14048, AW14049, AW14051, AW16052, AW16053, AW16054, AW16056, AW16057, AW16058, AW16059, AW16060, AW16061, AW20062, AW20063, AW20064, AW20065, AW20066, AW20067, AW20068, AW20069, AW20070, AW20071, AW23072, AW23073, AW23074, AW23075, AW23076, AW23077, AW23078, AW23079, AW28087, AW28088, AW28089, AW28090, AW28091, AW28094, AW28095, AW28096, AW28097, AW28098, AW31099, AW31100, AW31102, AW31103, AW31104, AW31105, AW31106, AW31107, AW31108, AW31109, AW31110, AW35111, AW35112, AW35113, AW35114, AW35115, AW35116, AW35117, AW35120, AW35121, AW35122, AW35123, AW35124, AW35125, AW35126, AW35127, AW35128, AW42129, AW42131, AW42132, AW42133, AW45134, AW45135, AW45136, AW45137, AW45138, AW45139, AW45140, AW45141, AW45143, AW45144, AW45145, AW45146, AW45147, AW47148, AW47149, AW47151, AW47152, AW47153, AW47154, AW47155, AW47156, AW47157, AW49159, AW49160, AW49161, AW49162, AW49163, AW49164, AW49165, AY01001, AY01002, AY01004, AY01005, AY01006, AY01007, AY01008, AY01009, AY01010, AY04011, AY04012, AY04013, AY04015, AY04016, AY04017, AY04018, AY04019, AY04020, AY08021, AY08022, AY08023, AY08024, AY08025, AY08026, AY08027, AY08028, AY08029, AY08032, AY08033, AY08034, AY10036, AY10037, AY10038, AY10039, AY10040, AY10041, AY10042, AY10044, AY10045, AY10046, AY10047, AY10048, AY10049, AY10051, AY10053, AY10054, AY10055, AY17056, AY17057, AY17058, AY17059, AY17060, AY17061, AY17062, AY17063, AY17064, AY17065, AY19066, AY19067, AY19069, AY19070, AY19071, AY19072, AY19073, AY19074, AY19075, AY23076, AY23077, AY23078, AY23079, AY23080, AY23081, AY30085, AY30086, AY30087, AY30088, AY30090, AY32092, AY32093, AY32094, AY32095, AY32096, AY32097, AY32098, AY32099, AY32100, AY32101, AY32102, AY32103, AY32104, AY32105, AY32106, AY35107, AY35108, AY35109, AY35110, AY35111, AY35112, AY35113, AY35114, AY35115, AY35116, AY35117, AY35118, AY35119, AY35120, AY35121, AY40123, AY40124, AY40125, AY40126, AY41127, AY41128, AY41129, AY41130, AY41132, AY41133, AY41134, AY41140, AY41141, AY51142, AY51143, AY51144, AY51145, AY51149, AY52150, AY52151, AY52152, AY52153, AY52154, AY52155, AY52156, AY52157, AY52158, AY52159, AZ02001, AZ02002, AZ02004, AZ02006, AZ02007, AZ02008, AZ02009, AZ02010, AZ06011, AZ06012, AZ06013, AZ06014, AZ06015, AZ06016, AZ06017, AZ06018, AZ06020, AZ06021, AZ06022, AZ08023, AZ08024, AZ08025, AZ08026, AZ08027, AZ08028, AZ08029, AZ08030, AZ08031, AZ12033, AZ12034, AZ12035, AZ12038, AZ12040, AZ12042, AZ12043, AZ12044, AZ17045, AZ17046, AZ17048, AZ17050, AZ17051, AZ17053, AZ18054, AZ18055, AZ18056, AZ18057, AZ19058, AZ19059, AZ19060, AZ19061, AZ20062, AZ20063, AZ20064, AZ20065, AZ20066, AZ20067, AZ20068, AZ20069, AZ20070, AZ22071, AZ22072, AZ22073, AZ22074, AZ22075, AZ23077, AZ23078, AZ23079, AZ23080, AZ24081, AZ24082, AZ24083, AZ24084, AZ24085, AZ25087, AZ25088, AZ25089, AZ25090, AZ28091, AZ28092, AZ28093, AZ28094, AZ28095, AZ28096, AZ28097, AZ28098, AZ28099, AZ28100, AZ34101, AZ34102, AZ34103, AZ34104, AZ34105, AZ34106, AZ34107, AZ34108, AZ34109, AZ34110, AZ34111, AZ34112, AZ34113, AZ34114, AZ34115, AZ39116, AZ39118, AZ39119, AZ39120, AZ39121, AZ39122, AZ39123, AZ39124, AZ39125, AZ41127, AZ41131, AZ41132, AZ41133, AZ41134, AZ41135, AZ43136, AZ43137, AZ43138, AZ43139, AZ43140, AZ43141, AZ43143, AZ46146, AZ46147, AZ46148, AZ46149, AZ46150, AZ46151, AZ46152, AZ46156, AZ46157, AZ46158, AZ46159, AZ48160, AZ48162, AZ48163, AZ48164, AZ48166, AZ48167, AZ49168, AZ49169, AZ49170, AZ49172, AZ49173, AZ49174, AZ49175, AZ49176, BA03001, BA03002, BA03003, BA03004, BA03005, BA03006, BA03007, BA03008, BA03009, BA03010, BA07011, BA07012, BA07013, BA07014, BA07015, BA07016, BA07017, BA07018, BA07019, BA07020, BA08021, BA08022, BA08023, BA08024, BA08025, BA10026, BA10027, BA10028, BA10029, BA10030, BA10031, BA13032, BA13034, BA13035, BA13036, BA13037, BA13038, BA13039, BA13040, BA13041, BA15042, BA15045, BA16047, BA16048, BA16049, BA16050, BA17052, BA17053, BA17054, BA18055, BA18056, BA18057, BA19060, BA19061, BA19062, BA21064, BA22067, BA22068, BA22069, BA22070, BA23071, BA23072, BA23073, BA24075, BA24076, BA24077, BA24078, BA25080, BA25081, BA25082, BA25083, BA25084, BA25085, BA27086, BA27087, BA27088, BA27089, BA27090, BA29091, BA29092, BA29093, BA29094, BA29095, BA31096, BA31097, BA31098, BA31099, BA31100, BA35101, BA35103, BA35104, BA35105, BA36106, BA36107, BA37109, BA37110, BA37111, BA37112, BA38113, BA38114, BA38115, BA40116, BA40117, BA40118, BA41119, BA41120, BA41121, BA41122, BA43124, BA43126, BA46127, BA46128, BA46129, BA46130, BA48131, BA48132, BA48133, BA49134, BA49135, BA49136, BA49137, BA50138, BA50139, BA50140, BA50141, BA50142, BB03001, BB03002, BB03003, BB03004, BB04005, BB04006, BB05007, BB05008, BB05009, BB06010, BB09011, BB09012, BB09013, BB09014, BB09015, BB16016, BB16017, BB16018, BB16019, BB16021, BB16022, BB16023, BB23024, BB23025, BB23026, BB23027, BB23028, BB27029, BB27030, BB27031, BB27032, BB27033, BB27034, BB47045, BB47046, BB47047, BB47048, BB47049, BB47050, BB47051, BB47052, BB47053, BB47054, BC01001, BC01002, BC01003, BC01004, BC04006, BC04007, BC04008, BC10016, BC10017, BC10018, BC10019, BC14021, BC14022, BC14023, BC14024, BC21026, BC21027, BC29030, BC29031, BC29032, BC29035, BC30036, BC30039.      Catalog Number B66705.     Serial Numbers:  AZ30002, AZ30003, AZ30004, AZ39005, BA14002.      Catalog Number C11477.  Serial Numbers:  BB11109, BB11110, BB12111, BB12117, BB12118, BB12119, BB12120, BB12121, BB12122, BB12123, BB18124, BB18125, BB18126, BB18127, BB18128, BB18129, BB18130, BB18133, BB23134, BB23135, BB23136, BB31145, BB31146, BB31147, BB31148, BB31149, BB31150, BB36157, BB36158, BB36159, BB36160, BB36161, BB36162, BB43163, BB43164, BB43165, BB43166, BB43167, BB43168, BB47169, BB47170, BB47171, BB47172, BB47173, BB47174, BB47176, BB47177, BC01101, BC01102, BC01103, BC01104, BC01105, BC04106, BC04108, BC04109, BC04110, BC04111, BC04112, BC04113, BC04114, BC04115, BC09117, BC09118, BC09119, BC09120, BC09121, BC09122, BC09123, BC09124, BC09125, BC09126, BC09127, BC09128, BC14131, BC14132, BC14133, BC14134, BC14135, BC14136, BC14137, BC14138, BC14140, BC14141, BC14142, BC14143, BC14144, BC14145, BC19146, BC19147, BC19148, BC19149, BC19150, BC19151, BC19152, BC19153, BC19154, BC19155, BC19156, BC19157, BC21158, BC21159, BC21160, BC21161, BC21162, BC21163, BC21164, BC21166, BC21169, BC21171, BC24172, BC24173, BC24174, BC24175, BC24176, BC24178, BC24180, BC24181, BC30182, BC30183, BC30184, BC30185, BC30186, BC30187, BC30188, BC30189, BC30190, BC30191, BC33192, BC33193, BC33194, BC33195, BC33196, BC33197, BC33198, BC33200, BC33201, BC33202, BC33203, BC33204, BC33206, BC36207, BC36208, BC36209, BC36210, BC36211, BC36212, BC36213, BC36214, BC36215.",
      "center_classification_date": "20191122",
      "distribution_pattern": "Worldwide distribution - US Nationwide distribution and countries of Australia, Bahrain, Belarus, Belgium, Brazil, Canada, China, Colombia, Czech Republic, Egypt, Ethiopia, France, Germany, Greece, Hong Kong, India, Israel, Italy, Japan, Kuwait, Latvia, Lebanon, Lithuania, Macau, Malaysia, Mayotte, Mexico, Morocco, Netherlands, New Zealand, Philippines, Portugal, Qatar, Russia, Saudi Arabia, Singapore, South Africa, South Korea, Spain, Switzerland, Taiwan, Thailand, Trinidad and Tobago, Turkey, United Arab Emirates, United Kingdom, Vietnam.",
      "state": "FL",
      "product_description": "UniCel DxH Slidemaker Stainer (DxH SMS) and UniCel DxH Slidemaker Stainer II (DxH SMS II) - Product Usage: The intended use is as follows: The DxH Slidemaker Stainer is a fully automated slide preparation and staining device that aspirates a whole-blood sample, smears a blood film on a clean microscope slide, and delivers a variety of fixatives, stains, buffers, and rinse solutions to that blood smear.",
      "report_date": "20191204",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Beckman Coulter, Inc.",
      "recall_number": "Z-0516-2020",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "product_type": "Devices",
      "event_id": "84186",
      "termination_date": "20210427",
      "more_code_info": "",
      "recall_initiation_date": "20191007",
      "postal_code": "33196-2500",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}