  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Bolton",
      "state": "N/A",
      "country": "Canada",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "81266",
      "recalling_firm": "PerkinElmer Health Sciences Canada, Inc.",
      "address_1": "32 Nixon Rd Unit 1",
      "address_2": "N/A",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "FL, MN, PA  India, Italy, Korea, Spain",
      "recall_number": "Z-0513-2019",
      "product_description": "PerkinElmer QSight 210 MD Mass Spectrometer Simplicity 3Q MD v1.0 with Hot Fix 2 Software intended to identify compounds in human samples by ionizing the compound and separating the resulting ions. Part Number: BC004872",
      "product_quantity": "18 units",
      "reason_for_recall": "Under certain conditions when using PerkinElmer QSight 210 MD mass spectrometer with Simplicity 3Q MD and Hotfix 2 software, it is possible for the original data from the last acquired sample to be over-written when the automatic flush is enabled.",
      "recall_initiation_date": "20180924",
      "center_classification_date": "20181128",
      "termination_date": "20190917",
      "report_date": "20181205",
      "code_info": "210MDBC709100008  210MD8C709140010  210MD8C709150009  210MDBC709160011  210MDBC709180012  210MDBC709200013  210MD8C710020014  210MDBC710050016  210MD8C710110018  210MDBC710170020  210MD8C710180022  210MD8C712220428  210MDBC801240030  210MDBC801250031  210MDBC802140032  210MD8C802150033  210MD8C802260035  210MDBC803190037"
    }
  ]
}