  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Natick",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "95601",
      "recalling_firm": "Zyno Medical LLC",
      "address_1": "177 Pine St",
      "address_2": "N/A",
      "postal_code": "01760-1331",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "This product is distributed only in the United States.",
      "recall_number": "Z-0506-2025",
      "product_description": "Z-800 Infusion System. Model/Catalog Number: Z-800, Z-800F, Z-800W, Z-800WF.      The Z-800 Infusion system is intended to provide intravenous infusion of parenteral fluids, blood, and blood products to a patient under the direction or supervision of a physician or other certified healthcare professional.",
      "product_quantity": "34,994 units",
      "reason_for_recall": "Potential for the battery to deplete over time from self-discharge and internal parasitic leakage current despite being continuously plugged into AC mains.",
      "recall_initiation_date": "20241018",
      "center_classification_date": "20241121",
      "report_date": "20241127",
      "code_info": "Model/Catalog Number: Z-800, Z-800F, Z-800W, Z-800WF  Lot Code: Lots: 201016280, 200515888, 200406416, 230920212, 200413400, 20040660, 200716968, 201016720, 20071632, 201211740, 201208256, 210319640, 210709900, 201211360, 2101221000, 210913560, 210913200, 231008352, 220205400, 220204548, 2211141000, 230206980, 230620800, 230420548, 230420136, 230525452, 230620200."
    }
  ]
}