  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Guilford",
      "state": "CT",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "95594",
      "recalling_firm": "Defibtech, LLC",
      "address_1": "741 Boston Post Rd",
      "address_2": "Suite 201",
      "postal_code": "06437-2714",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "Ireland",
      "recall_number": "Z-0504-2025",
      "product_description": "Defibtech DDP-2001: Adult Defibrillation Pads  For use with AED Models: Lifeline View Auto, lifeline Pro, lifeline ECG, lifeline View (DDU-2XXX Series). -Indicated for use on victims of sudden cardiac arrest (SCA.)",
      "product_quantity": "22,294 total",
      "reason_for_recall": "Within the Irish Market, that a Field Safety Notice is being issued due to a select number of authorized labelled Defibtech AED Pads, which have had the authorized label removed and an unauthorized  label applied,  could potentially lead to expired Defibtech AED Pads appearing as though they are within the acceptable shelf life.there is a possibility of hydrogel drying, which may result in poor  electrical connection to the patient, which can reduce the energy delivered to the patient",
      "recall_initiation_date": "20241014",
      "center_classification_date": "20241121",
      "report_date": "20241204",
      "code_info": "UDI-DI: 00815098020447, DDP-2001 adult defibrillation pads, single pack; 30815098020448, DDP-2001 adult defibrillation pads, 10 pairs; 50815098020442, DDP-2001 adult defibrillation pads, 40 pairs",
      "more_code_info": ""
    }
  ]
}