  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Oxford",
      "state": "N/A",
      "country": "United Kingdom",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "95581",
      "recalling_firm": "OrganOx Ltd",
      "address_1": "Magdalen Centre",
      "address_2": "1 Robert Robinson Avenue",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "California, Texas, Ohio, Florida, New York, Massachusetts, Virginia, Oklahoma, DC, South Carolina and Georgia.",
      "recall_number": "Z-0501-2025",
      "product_description": "OrganOx metra -Intended as a Normothermic Machine Perfusion system of donor transport livers.    Model/Catalog Number: D0003    The OrganOx metra is a transportable device intended to be used to sustain donor livers destined for transplantation in a functioning state for a total preservation time of up to 12 hours.",
      "product_quantity": "20 units",
      "reason_for_recall": "Battery movement led to damage of one of the battery cells causing it to short to the battery retention bracket",
      "recall_initiation_date": "20241015",
      "center_classification_date": "20241121",
      "report_date": "20241127",
      "code_info": "UDI-DI:  05060462240029  Serial numbers 511, 512, 514, 515, 516, 517, 518, 519, 520, 522, 523, 524, 525, 527, 528, 529, 530, 531, 533, 535, 536, 537"
    }
  ]
}