{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Franklin Lakes",
      "state": "NJ",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "81148",
      "recalling_firm": "Becton Dickinson & Company",
      "address_1": "1 Becton Dr",
      "address_2": "N/A",
      "postal_code": "07417-1815",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "Illinois",
      "recall_number": "Z-0493-2019",
      "product_description": "KIT URIN CUP PLC 16X100 10.0 UA YEL, Cat. No. 364981: Urine Collection Kit for collection and transport of urine specimens for urinalysis",
      "product_quantity": "Total: 29,982,150 BD cups (US); 97,800,900 BD cups (OUS)",
      "reason_for_recall": "BD Vacutainer Urine Collection Cups may contribute to a slight elevation of Red Blood Cell (RBC) counts when used with the Sysmex UF Series Analyzers. BD recommends following Good Clinical Practice (GCP) by using microscopy to confirm an elevated RBC Count in a Urine sample as per CLSI GP16-3A. 29(4) and CAP Checklist on Urinalysis. Product will continue to ship to not impact critical patient care.",
      "recall_initiation_date": "20180830",
      "center_classification_date": "20181121",
      "termination_date": "20201021",
      "report_date": "20181128",
      "code_info": "All lots within expiry",
      "more_code_info": ""
    }
  ]
}