  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "East Walpole",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "97819",
      "recalling_firm": "Siemens Healthcare Diagnostics, Inc.",
      "address_1": "333 Coney St",
      "address_2": "",
      "postal_code": "02032-1516",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide - US Nationwide distribution in the state of CA and the countries of Argentina, Australia, Austria, Belgium, Bosnia Herzeg. , Brazil, Canary Islands, Chile, Colombia, Costa Rica, Croatia, Czech Republic, Estonia, France, Germany, Greece, Guatemala, Hong Kong, Hungary, Israel, Italy, Latvia, Macedonia, Mexico, Netherlands, Norway, Pakistan, Panama, Poland, Portugal, Romania, Russian Fed., Serbia, Slovakia, Spain, Sweden, Turkey, U.A.E., Ukraine, United Kingdom, Uruguay.",
      "recall_number": "Z-0491-2026",
      "product_description": "IMMULITE 2000 Oak Mix Specific Allergen;  Siemens Material Number (SMN): 10386105;  Catalog Number: L2KUN6;",
      "product_quantity": "34 units (1 US, 33 OUS)",
      "reason_for_recall": "The barcode orientation on affected devices causes incorrect scanning order of tubes within the allergen holder wedge. When using an IMMULITE 2000 / IMMULITE 2000 XPi instrument in combination with a Zebra barcode scanner, the instrument will not correctly recognize the position of tubes within the allergen holder wedge.",
      "recall_initiation_date": "20251006",
      "center_classification_date": "20251113",
      "report_date": "20251119",
      "code_info": "Siemens Material Number (SMN): 10386105; Model/Catalog: T77L2; UDI-DI: 00630414965970; Lot number: 627;",
      "more_code_info": ""
    }
  ]
}