  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Indianapolis",
      "state": "IN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "95609",
      "recalling_firm": "Roche Diagnostics Operations, Inc.",
      "address_1": "9115 Hague Rd",
      "address_2": "N/A",
      "postal_code": "46256-1025",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US:  NJ, TX, NY, FL, IN, TX, OK, SC, IA,",
      "recall_number": "Z-0489-2025",
      "product_description": "cobas HCYS, Homocysteine Enzymatic Assay, Material Number REF 06542921190, for use on cobas c 701/702, in vitro test for total L-homocysteine",
      "product_quantity": "458 units",
      "reason_for_recall": "Abnormal calibration signals and control recovery issues on the Homocysteine Enzymatic (HCYS) Assay can produce erroneous patient results (positive or negative bias)",
      "recall_initiation_date": "20241017",
      "center_classification_date": "20241119",
      "report_date": "20241127",
      "code_info": "UDI/DI 04015630929979, Lot Number 747029, exp. 30-SEPT-2024"
    }
  ]
}