      "status": "Ongoing",
      "city": "deerfield",
      "state": "IL",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "91200",
      "recalling_firm": "Leica Microsystems, Inc.",
      "address_1": "10 Parkway North blvd ste 300",
      "address_2": "N/A",
      "postal_code": "60089-6622",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide",
      "recall_number": "Z-0477-2023",
      "product_description": "Leica MICROSYSTEMS M530 OHX, Part Number 10448737, surgical microscope system",
      "product_quantity": "148 units",
      "reason_for_recall": "There is a potential issue resulting in the Power Supply potentially becoming overheated due to internal components overstressing. Consequently, due to the overstressing of the internal components, one specific component can fail which will cause the device to unexpectedly shutdown.",
      "recall_initiation_date": "20221121",
      "center_classification_date": "20221209",
      "report_date": "20221221",
      "code_info": "Serial Numbers: 10622001, 10622002, 10822001, 20322002, 20522001, 20622001, 30122001, 30322001, 30522001, 30622002, 30722001, 31221001, 40422001, 41221001, 50122001, 60122001, 60522001, 60622001, 70122001, 70322001, 70422001, 70622001, 80422001, 80422002, 80722001, 80822001, 90122001, 90622001, 91221001, 100122001, 100122002, 100322002, 100622001, 100622002, 110222001, 110522001, 110822001, 111221001, 120222001, 120522001, 130122001, 130522001, 130622001, 140122001, 140222001, 140422001, 140522001, 141221001, 151221001, 160122001, 160322001, 160622001, 160622002, 170222001, 170522001, 190322001, 190622001, 200122001, 200522001, 200722001, 210422001, 210622001, 230222001, 230622001, 240122001, 240522001, 260522001, 260722001, 260722002, 261121001, 261221001, 270122001, 270422001, 270422002, 270522001, 270622001, 271121001, 280222002, 280622001, 290522001, 290622001, 291221001, 300322001, 300622002, 301121001, 301221001, 310322001, 310522001.    ***EXPANSION 09/08/2023, Serial Numbers: 220523002, 220822001, 131222001, 251122001, 271122001, 100223001, 220423001, 300822001, 30323001, 41222001, 51222001, 91222002, 81222002, 301122001, 180822001, 71222001, 180722001, 220523001, 300523002, 80623001, 30523001, 11122001, 31122001, 81122001, 91222001, 30723001, 211122001, 121222001, 310523001, 40423001, 290323001, 300323001, 141222001, 141222002, 310822001, 40623002, 300922001, 200223001, 90223001, 180523001, 280523001, 240423001, 10623001, 191122001, 61222001, 90623001, 190423001, 70223001, 90223002, 40723001, 130623001, 180423001, 230922001, 260922001, 251022001, 291022001, 200323001, 261022001, 140523001, 160523001."
    }
  ]
}