  "results": [
    {
      "status": "Ongoing",
      "city": "Tempe",
      "state": "AZ",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "95491",
      "recalling_firm": "Bard Peripheral Vascular Inc",
      "address_1": "1625 W 3rd St Bldg 1",
      "address_2": "850 W. Rio Salado Prkwy.",
      "postal_code": "85281-2438",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide - US Nationwide including in the states of WA, LA, MN, PA, UT, SC, OK, KY, AZ, IL, TX, NJ, NY, FL, TN, IN, OH, NC, CO, VA, GA, MD, CA, AL, WI, MO, IA, CT, NE, MI, MS, NV, KS, OR, SD, ID, NH, RI, WY, AR, MA, DE, ND, NM, AK, PR and the countries of Portugal, Spain, South Africa, Greece, United Arab Emirates, United Kingdom, Sweden, Finland, Belgium, Korea, Australia, Canada.",
      "recall_number": "Z-0467-2025",
      "product_description": "Bard Marquee Disposable Core Biopsy Instrument Kit:  BARD MARQUEE KIT 12G X 10 CM, REF: MQK1210;  BARD MARQUEE 12G X 10 CM, REF: MQ1210;  BARD MARQUEE Kit 12G X 13 CM, REF: MQK1213;  BARD MARQUEE 12G X 13 CM, REF: MQ1213",
      "product_quantity": "71,498",
      "reason_for_recall": "Disposable biopsy instruments were potentially manufactured missing the backend notch retention feature on the needle that results in inadequate securing of the needle inside the device, which could lead to needle detachment and misplacement, which could lead to tissue damage and bleeding.",
      "recall_initiation_date": "20240930",
      "center_classification_date": "20241115",
      "report_date": "20241127",
      "code_info": "REF/UDI-DI/Lot(Expiration): MQK1210/00801741097263/1490758(9/28/2025), 1506044(1/28/2026), 1506045(1/28/2026), 1511827(2/28/2026), 1514692(3/28/2026), 1518977(4/28/2026), 1518978(4/28/2026), 1518979(4/28/2026), 1519143(4/28/2026), 1520507(5/28/2026), 1520508(5/28/2026), 1522157(5/28/2026), 1522158(5/28/2026), 1522577(5/28/2026), 1523785(6/28/2026), 1523914(6/28/2026), 1523915(6/28/2026), 1525370(6/28/2026), 1525371(6/28/2026), 1525373(6/28/2026), 1528417(7/28/2026), 1528418(7/28/2026), 1528495(7/28/2026), 1528496(7/28/2026), 1528497(7/28/2026), 1528560(7/28/2026), 0001618867(3/28/2028), 0001620281(3/28/2028), 0001632675(6/28/2028);  MQ1210/00801741097119/1514460(3/28/2026), 1514461(3/28/2026), 1519144(4/28/2026), 1519145(4/28/2026), 1519456(4/28/2026), 1520066(5/28/2026), 1520067(5/28/2026), 1520506(5/28/2026), 1523783(6/28/2026), 1523784(6/28/2026), 0001618258(3/28/2028), 0001631611(6/28/2028), 0001633216(7/28/2028);  MQK1213/00801741097270/1543517(10/28/2026), 1559312(2/28/2027), 1562724(3/28/2027), 1569737(4/28/2027), 1581652(7/28/2027);  MQ1213/00801741097126/1558191(2/28/2027), 1559313(2/28/2027)"
    }
  ]
}